FDA Adverse Event Injury Summary report: N

WOUND DRAIN

MDR report key: 3142482 · Received May 8, 2013

Report

Report Number
1018233-2013-01821
Event Type
Injury
Date Received
May 8, 2013
Report Date
April 12, 2013
Manufacturer
PRODUCTOS PARA EL CUIDADO DE LA SALUD
Product Code
GBX
PMA / PMN Number
EXEMPT
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE SAMPLE WAS NOT RETURNED. THE FINISHED PRODUCT MET ALL SPECIFICATIONS PRIOR TO BEING RELEASED FOR GENERAL DISTRIBUTION. INSTRUCTIONS FOR USE PER (B)(4) STATE THE FOLLOWING PRECAUTIONS TO AVOID THE POSSIBILITY OF DRAIN DAMAGE OR BREAKAGE: AVOID SUTURING THROUGH DRAINS. DRAINS SHOULD LIE FLAT AND IN LINE WITH THE SKIN EXIT AREAS. PARTICULAR CARE SHOULD BE TAKEN TO AVOID ANY OBSTACLES TO THE DRAIN EXIT PATH. DRAINS SHOULD BE CHECKED FOR FREE MOTION DURING CLOSURE TO MINIMIZE THE POSSIBILITY OF BREAKAGE. DRAIN REMOVAL SHOULD BE DONE GENTLY BY HAND. DRAINS SHOULD NOT BE HANDLED WITH POINTED, TOOTHED OR SHARP INSTRUMENTS WHICH COULD CAUSE CUTS OR NICKS AND LEAD TO SUBSEQUENT STRUCTURAL FAILURE OF THE DRAIN. SURGICAL REMOVAL MAY BE NECESSARY IF DRAIN IS DIFFICULT TO REMOVE OR BREAKS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DRAIN BROKE OFF IN THE PT AND THE PT WAS RETURNED TO SURGERY ON (B)(6) 2013 TO REMOVE THE PIECE OF BROKEN DRAIN. IT WAS REPORTED THAT THE PT IS DOING FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
202650 WOUND DRAIN GBX PRODUCTOS PARA EL CUIDADO DE LA SALUD NA UNK

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention