DIMENSION(R) CLINICAL CHEMISTRY SYSTEM
Report
- Report Number
- 2517506-2013-00243
- Event Type
- Malfunction
- Date Received
- June 3, 2013
- Date of Event
- April 10, 2013
- Report Date
- May 7, 2013
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
- Product Code
- LCP
- PMA / PMN Number
- K963498
- Removal / Correction Number
- 2517506-05-17-2013-006-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AR, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
SIEMENS HEALTHCARE DIAGNOSTICS HAS CONFIRMED COMPLAINTS OF DIMENSION(R) CHEMISTRY WASH CAUSING QC AND PATIENT SAMPLE RESULT SHIFTS. INTERNAL INVESTIGATION HAS CONFIRMED SHIFTS CAN BE ENCOUNTERED IN TWO SCENARIOS: WHEN AN AFFECTED LOT OF CHEMISTRY WASH IS PLACED ON THE INSTRUMENT FOLLOWING AN UNAFFECTED LOT, A NEGATIVE BIAS WILL BE SEEN ON CTNI, LTNI, TSH, PBNP AND LPBN. A POSITIVE BIAS WILL BE SEEN ON FT4. IF ANY OF THESE ASSAYS ARE CALIBRATED USING AN AFFECTED LOT OF CHEMISTRY WASH, QC SHOULD RETURN WITHIN RANGE; BUT WHEN A SUBSEQUENT UNAFFECTED CHEMISTRY WASH LOT IS PLACED INTO USE, A POSITIVE BIAS WILL BE SEEN ON CTNI, LTNI, TSH, PBNP AND LPBN. A NEGATIVE BIAS WILL BE SEEN ON FT4. SIEMENS HEALTHCARE DIAGNOSTICS INC. ISSUED AN URGENT MEDICAL DEVICE CORRECTION LETTER, (B)(4), IN (B)(4) 2013. THE LETTER STATED THAT SIEMENS HAS CONFIRMED THAT CHANGING THE CHEMISTRY WASH LOT CAN CAUSE QC AND PATIENT SAMPLE RESULT SHIFTS. THE LETTER DIRECTED CUSTOMERS TO DISCONTINUE USE AND DISCARD REMAINING INVENTORY OF THE FOLLOWING LOTS OF CHEMISTRY WASH: RD23031, RD23111, RD23241, RD23311, RD23391, AND RD23461. THE LETTER ADVISED CUSTOMERS TO ENSURE CTNI, LTNI, TSH, FT4, PBNP AND/OR LPBN ASSAYS HAVE BEEN CALIBRATED USING AN UNAFFECTED LOT OF CHEMISTRY WASH.
A CUSTOMER COMPLAINS OF QC SHIFTS WITH FREE THYROXINE (FT4) WITH DIMENSION(R) CHEMISTRY WASH LOT RD23311. PATIENT RESULTS WERE NOT REPORTED TO THE PHYSICIANS WHILE QC WAS OUT OF RANGE. PATIENT TREATMENT WAS NOT ALTERED OR PRESCRIBED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE SHIFTS IN FT4 RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 244137 | DIMENSION(R) CLINICAL CHEMISTRY SYSTEM | CHEMISTRY WASH | LCP | SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW | RD23311 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |