ONCOR IMPRESSION PLUS
Report
- Report Number
- 2910081-2013-03845
- Event Type
- Other
- Date Received
- May 28, 2013
- Date of Event
- April 25, 2013
- Report Date
- April 26, 2013
- Manufacturer
- SIEMENS MEDICAL SOLUTIONS USA, INC. - RADIATION ONCOLOGY
- Product Code
- IYE
- PMA / PMN Number
- K123812
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
SIEMEN'S INVESTIGATION INTO THE REPORTED OCCURRENCE REVEALS THAT THE TABLE MOVEMENT HAPPENS DUE TO WORKFLOW ERRORS THAT OCCURS WHEN VARIAN TREATMENT IS CONNECTED TO SIEMENS ONCOR LINAC SYSTEMS. HOWEVER, SIEMENS LINAC SYSTEM OPERATED AS SPECIFIED ACCORDING TO THE TREATMENT INFORMATION RECEIVED FROM VARIAN TREATMENT AND (B)(4) 2.3.33. THE REPORTED INFORMATION HAS BEEN FORWARDED TO VARIAN FOR INVESTIGATION.
SIEMENS WAS INFORMED ON (B)(4) 2013 THAT THE VARIAN TREATMENT INTERFACE USED WITH RT THERAPIST (B)(4) 2.3.33 IN CONNECTION WITH THE ONCOR LINAC SYSTEM ALLOWED TREATMENT TABLE MOVEMENT AFTER THE FIRST SEGMENT FIELD WAS TREATED. REPORTEDLY, THE TREATMENT TABLE MOVEMENTS OCCURRED DURING AUTO FIELD SEQUENCING FOR THREE PTS THAT RESULTED IN MISTREATMENTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 232891 | ONCOR IMPRESSION PLUS | ACCELERATOR, LINEAR, MEDICAL | IYE | SIEMENS MEDICAL SOLUTIONS USA, INC. - RADIATION ONCOLOGY | 05857912 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | RT THERAPIST ((B)(4)) 2.3.33| VARIAN TREATMENT| ONCOR: SW VERSION 9.2.18 |