FDA Adverse Event Other Summary report: N

ONCOR IMPRESSION PLUS

MDR report key: 3142285 · Received May 28, 2013

Report

Report Number
2910081-2013-03845
Event Type
Other
Date Received
May 28, 2013
Date of Event
April 25, 2013
Report Date
April 26, 2013
Manufacturer
SIEMENS MEDICAL SOLUTIONS USA, INC. - RADIATION ONCOLOGY
Product Code
IYE
PMA / PMN Number
K123812
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SIEMEN'S INVESTIGATION INTO THE REPORTED OCCURRENCE REVEALS THAT THE TABLE MOVEMENT HAPPENS DUE TO WORKFLOW ERRORS THAT OCCURS WHEN VARIAN TREATMENT IS CONNECTED TO SIEMENS ONCOR LINAC SYSTEMS. HOWEVER, SIEMENS LINAC SYSTEM OPERATED AS SPECIFIED ACCORDING TO THE TREATMENT INFORMATION RECEIVED FROM VARIAN TREATMENT AND (B)(4) 2.3.33. THE REPORTED INFORMATION HAS BEEN FORWARDED TO VARIAN FOR INVESTIGATION.

Description of Event or Problem · 1

SIEMENS WAS INFORMED ON (B)(4) 2013 THAT THE VARIAN TREATMENT INTERFACE USED WITH RT THERAPIST (B)(4) 2.3.33 IN CONNECTION WITH THE ONCOR LINAC SYSTEM ALLOWED TREATMENT TABLE MOVEMENT AFTER THE FIRST SEGMENT FIELD WAS TREATED. REPORTEDLY, THE TREATMENT TABLE MOVEMENTS OCCURRED DURING AUTO FIELD SEQUENCING FOR THREE PTS THAT RESULTED IN MISTREATMENTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
232891 ONCOR IMPRESSION PLUS ACCELERATOR, LINEAR, MEDICAL IYE SIEMENS MEDICAL SOLUTIONS USA, INC. - RADIATION ONCOLOGY 05857912 NA

Patients

Seq Age Sex Outcome Treatment
1 RT THERAPIST ((B)(4)) 2.3.33| VARIAN TREATMENT| ONCOR: SW VERSION 9.2.18