FDA Adverse Event Malfunction Summary report: N

GYNECARE MORCELLEX* TISSUE MORCELLATOR

MDR report key: 3142264 · Received June 3, 2013

Report

Report Number
2210968-2013-06375
Event Type
Malfunction
Date Received
June 3, 2013
Date of Event
May 10, 2013
Report Date
May 10, 2013
Manufacturer
ETHICON, INC.
Product Code
HET
PMA / PMN Number
K100280
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4): THE ACTUAL DEVICE INVOLVED IN THIS EVENT WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY AND FUNCTIONALLY EVALUATED. THE BLADE FAILED TO ROTATE/ADVANCE DURING THE EVALUATION; IT IS LIKELY THAT THE ACCUMULATION OF BLOOD, BODY FLUIDS, AND TISSUE PREVENTED THE DEVICE FROM CONTINUING TO OPERATE AS INTENDED.

Additional Manufacturer Narrative · 1

(B)(4): THE PRODUCT UPON WHICH THIS MEDWATCH IS BASED IS ANTICIPATED. ONCE THE PRODUCT IS RECEIVED, ANY FURTHER INFORMATION DERIVED FROM THE EVALUATION WILL BE SUBMITTED IN A SUPPLEMENTAL 3500A FORM.

Additional Manufacturer Narrative · 1

(B)(4). IN ADDITION, A REVIEW OF THE BATCH MANUFACTURING RECORDS WAS CONDUCTED AND THE BATCH MET ALL FINISHED GOODS RELEASE CRITERIA.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PATIENT UNDERWENT A ROBOTIC HYSTERECTOMY WITH BILATERAL SALPINGO-OOPHORECTOMY AND CYSTOSCOPY ON (B)(4) 2013. DURING THE PROCEDURE, THE DEVICE BLADE WOULD NOT EXTEND OR RETRACT. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH NO ADVERSE PATIENT CONSEQUENCES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
243098 GYNECARE MORCELLEX* TISSUE MORCELLATOR LAPAROSCOPE, GYNECOLOGIC HET ETHICON, INC. NA MT217855

Patients

Seq Age Sex Outcome Treatment
1