FDA Adverse Event Malfunction Summary report: N

FRONT CUTTER

MDR report key: 3142263 · Received June 3, 2013

Report

Report Number
8030965-2013-02628
Event Type
Malfunction
Date Received
June 3, 2013
Report Date
May 6, 2011
Manufacturer
SYNTHES GMBH
Product Code
HXZ
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO FDA WARNING LETTER DATED FEBRUARY 2012. DEVICE(S) LISTED IN THIS REPORT IS (ARE) USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. THE VISUAL INSPECTION OF THE FRONT CUTTER REVEALED THAT ONE OF THE CUTTING JAWS BROKE OFF AND THAT THE CUTTING FACES SHOWED SIGNS OF CONSIDERABLE WEAR AND TEAR. UNFORTUNATELY, THE COURSE OF EVENTS LEADING TO THE DAMAGE CANNOT BE ESTABLISHED WITHOUT FURTHER DETAILED INFORMATION. THE INSTRUMENT IN QUESTION IS ALREADY FOUR YEARS OLD AND HAS BEEN INTENSELY USED OVER AN EXTENDED PERIOD OF TIME. WE BELIEVE THAT SEVERAL INSTANCES OF EXCESSIVE STRESS OVER AND ABOVE THE LIMITS OF THE MATERIAL MAY HAVE LED TO BREAKAGE/FATIGUE. THE SURFACE ALONG THE BREAKAGE IS HOMOGENOUS INDICATING THAT THE MATERIAL QUALITY WAS FLAWLESS. INSPECTION IN ACCORDANCE WITH PRODUCTION AND MATERIAL DOCUMENTS INDICATED THAT THE HEAD CUTTER IS ENTIRELY COMPATIBLE WITH THE SPECIFICATIONS

Additional Manufacturer Narrative · 1

SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO FDA WARNING LETTER DATED FEBRUARY 2012. BLANK FIELDS ON THIS FORM INDICATE THE INFORMATION IS UNKNOWN, UNAVAILABLE OR UNCHANGED. DEVICE(S) LISTED IN THIS REPORT IS (ARE) USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. (B)(6). MANUFACTURING DOCUMENTS WERE REVIEWED AND NO COMPLAINT RELATED ISSUES WERE FOUND.

Description of Event or Problem · 1

CUTTING JAW OF CABLE CUTTER INSTRUMENT BROKE. THIS IS 1 OF 1 REPORT FOR THIS COMPLAINT (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
243210 FRONT CUTTER HXZ SYNTHES GMBH 1518878

Patients

Seq Age Sex Outcome Treatment
1