FDA Adverse Event Injury Summary report: N

CHARGING SYSTEM

MDR report key: 3142233 · Received May 29, 2013

Report

Report Number
1627487-2013-15698
Event Type
Injury
Date Received
May 29, 2013
Date of Event
May 6, 2013
Report Date
May 6, 2013
Manufacturer
ST. JUDE MEDICAL - NEUROMODULATION
Product Code
LGW
PMA / PMN Number
P010032
Removal / Correction Number
1627487-07262012-001-C
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THIS CHARGER MODEL WAS ASSOCIATED WITH A FIELD CORRECTION. MFR'S EVAL: CORRECTION AND PREVENTIVE ACTION (CAPA) INVESTIGATION WAS PERFORMED. EVAL CODES, RESULTS: POCKET HEATING WAS CONFIRMED. THE INVESTIGATION FOR CAPA (B)(4) ASSOCIATED WITH HEATING WHILE CHARGING (POCKET HEATING) CONCLUDED THAT THE CHARGER WAS CAPABLE OF TRANSFERRING ENERGY TO THE IPG AT A RATE THAT WOULD CAUSE HEATING OF THE IPG AND/OR CHARGING WAND OF SUFFICIENT ELEVATED TEMPERATURE TO CAUSE PAIN AND BURNS. THE HEATING WHILE CHARGING WAS DETERMINED TO BE EXACERBATED BY OFF-AXIS CHARGING OF SHALLOW IMPLANTED IPGS AND THAT ALL CHARGERS WERE CAPABLE OF ELEVATED HEATING. SJM HAS LIMITED INFO RELATED TO THE PT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Description of Event or Problem · 1

DEVICE 2 OF 2. REFERENCE MFR REPORT: 1627487-2013-15697. THE PT REPORTED HE IS EXPERIENCING UNCOMFORTABLE HEATING AT THE IPG SITE WHILE CHARGING. PT INDICATED THE HEATING IS UNBEARABLE AND BEGINS WITHIN 5-10 MINUTES AFTER CHARGING BEGINS. THE PT WAS SENT A REPLACEMENT LE CHARGING SYSTEM. F/U IDENTIFIED THE PT'S ISSUE WAS RESOLVED WITH THE USE OF THE REPLACEMENT CHARGING SYSTEM. ON 08/01/2012 ST, JUDE MEDICAL, NEUROMODULATION DIVISION, SENT FIELD ACTION LETTERS TO PTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS WAS EXPECTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
237485 CHARGING SYSTEM SCS CHARGING SYSTEM LGW ST. JUDE MEDICAL - NEUROMODULATION 3721 3835085

Patients

Seq Age Sex Outcome Treatment
1 46 YR Required Intervention SCS ANCHOR: MODEL 1194 (S)| IMPLANT DATE:| SCS LEAD: MODEL 3228| IMPLANT DATE: