FDA Adverse Event Injury Summary report: N

CHARGING SYSTEM

MDR report key: 3142149 · Received May 29, 2013

Report

Report Number
1627487-2013-06367
Event Type
Injury
Date Received
May 29, 2013
Date of Event
May 8, 2013
Report Date
May 8, 2013
Manufacturer
ST. JUDE MEDICAL - NEUROMODULATION
Product Code
LGW
PMA / PMN Number
P010032
Removal / Correction Number
1627487-07262012-001-C
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WV, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THIS CHARGER MODEL WAS ASSOCIATED WITH A FIELD CORRECTION. CORRECTION AND PREVENTIVE ACTION (CAPA) INVESTIGATION WAS PERFORMED. EVAL RESULTS: POCKET HEATING WAS CONFIRMED. THE INVESTIGATION FOR CAPA (B)(4) ASSOCIATED WITH HEATING WHILE CHARGING (POCKET HEATING) CONCLUDED THAT THE CHARGER WAS CAPABLE OF TRANSFERRING ENERGY TO THE IPG AT A RATE THAT WOULD CAUSE HEATING OF THE IPG AND/OR CHARGING WAND OF SUFFICIENT ELEVATED TEMPERATURE TO CAUSE PAIN AND BURNS. THE HEATING WHILE CHARGING WAS DETERMINED TO BE EXACERBATED BY OFF-AXIS CHARGING OF SHALLOW IMPLANTED IPGS AND THAT ALL CHARGERS WERE CAPABLE OF ELEVATED HEATING. SJM HAS LIMITED INFORMATION RELATED TO THE PT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Description of Event or Problem · 1

DEVICE 2 OF 2. REFERENCE MFR REPORT: 1627487-2013-06366. IT WAS REPORTED THE PT IS EXPERIENCING HEATING AT HIS IPG POCKET WHILE CHARGING. THE PT STATED THE HEATING IS UNCOMFORTABLE ENOUGH THAT HE HAS TO STOP CHARGING AND WAIT FOR THE AREA TO COOL OFF BEFORE RESUMING CHARGING. A NEW LOW ENERGY CHARGER WAS SENT TO ADDRESS THE ISSUE. ON 08/01/2012, ST. JUDE MEDICAL, NEUROMODULATION DIVISION, SENT A FIELD ACTION LETTERS TO PTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PTS. AN INCREASED IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS WAS EXPECTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
235210 CHARGING SYSTEM SCS CHARGING SYSTEM LGW ST. JUDE MEDICAL - NEUROMODULATION 3721 3593372

Patients

Seq Age Sex Outcome Treatment
1 80 YR Required Intervention SCS ANCHOR: MODEL 1192 (X2)| IMPLANT DATE:| SCS LEAD: MODEL 3186 (X2)| IMPLANT DATE: