CHARGING SYSTEM
Report
- Report Number
- 1627487-2013-04652
- Event Type
- Injury
- Date Received
- May 28, 2013
- Date of Event
- May 5, 2013
- Report Date
- May 5, 2013
- Manufacturer
- ST JUDE MEDICAL - NEUROMODULATION
- Product Code
- LGW
- PMA / PMN Number
- P010032
- Removal / Correction Number
- 1627487-07262012-001-C
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THIS CHARGER MODEL WAS ASSOCIATED WITH A FIELD CORRECTION. CORRECTIVE AND PREVENTIVE ACTION (CAPA) INVESTIGATION WAS PERFORMED. RESULTS - POCKET HEATING WAS CONFIRMED. THE INVESTIGATION FOR (B)(4) ASSOCIATED WITH HEATING WHILE CHARGING (POCKET HEATING) CONCLUDED THAT THE CHARGER WAS CAPABLE OF TRANSFERRING ENERGY TO THE IPG AT A RATE THAT WOULD CAUSE HEATING OF THE IPG AND/OR CHARGING WAND OF SUFFICIENT ELEVATED TEMPERATURE TO CAUSE PAIN AND BURNS. THE HEATING WHILE CHARGING WAS DETERMINED TO BE EXACERBATED BY OFF-AXIS CHARGING OF SHALLOW IMPLANTED IPGS AND THAT ALL CHARGES WERE CAPABLE OF ELEVATED HEATING. SJM HAS LIMITED INFO RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.
DEVICE 2 OF 2. REF MFR REPORT: 1627487-2013-04651. IT WAS REPORTED THE PT HAD EXPERIENCED POCKET HEATING WHILE USING THE CHARING SYSTEM. THE SJM REP MET WITH THE PT AND PROVIDED A REPLACEMENT CHARGING SYSTEM. IT WAS REPORTED THE PT HAD BEEN USING THE REPLACEMENT CHARGING SYSTEM, AND NO FURTHER HEATING HAD BEEN EXPERIENCED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 233243 | CHARGING SYSTEM | SCS CHARGING SYSTEM | LGW | ST JUDE MEDICAL - NEUROMODULATION | 3721 | 3577897 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 80 YR | Required Intervention | SCS LEAD: MODEL 3186 (2)| IMPLANT DATE:| IMPLANT DATE:| SCS ANCHOR: MODEL 1192 (2) |