LONG NAIL KIT R1.5, TI, RIGHT GAMMA3 Ø10X360MM X 125°
Report
- Report Number
- 0009610622-2013-00308
- Event Type
- Injury
- Date Received
- June 3, 2013
- Date of Event
- May 1, 2013
- Report Date
- May 13, 2013
- Manufacturer
- STRYKER OSTEOSYNTHESIS-KIEL
- Product Code
- HSB
- PMA / PMN Number
- K034002
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- PHYSICIAN
Narratives
ONCE THE INVESTIGATION HAS BEEN COMPLETED ANY ADDITIONAL INFORMATION WILL BE REPORTED IN A SUPPLEMENTAL REPORT.
EVALUATION SUMMARY: THE RETURNED DEVICE MATCHED THE REPORT, AND THE INCIDENT WAS CONFIRMED. INVESTIGATION REVEALED NO DISCREPANCIES IN MATERIAL OF MANUFACTURING. ACCORDING TO THE DAMAGE AND THE EVIDENCES OF THE BREAKAGE SURFACES THE NAIL BROKE IN A FATIGUE FRACTURE DUE TO OVERLOAD, CONTRIBUTED BY INTRA OPERATIVELY DAMAGE OF THE NAIL CAUSED BY A DEVIATED LAG SCREW STEP DRILL. DUE TO A NOTCHING EFFECT, THE MISS DRILLING MOST LIKELY HAD CREATED THE STARTING POINT OF THE IMPLANT BREAKAGE AT THE LATERAL EDGE OF THE ANTERIOR BRIDGE. WE DID NOT RECEIVE A REASONABLE STATEMENT REGARDING POSTOPERATIVE MOBILIZATION INSTRUCTIONS / WEIGHT BEARING. NEVERTHELESS, AS THE PATIENT WAS REVISED IT CAN BE ASSUMED THAT BONE CONSOLIDATION WAS INSUFFICIENT. CONSEQUENTLY, THE NAIL WAS NOT RELIEVED BY BONE CONSOLIDATION AND HAD TO ABSORB THE COMPLETE LOAD APPLICATION DURING THE IMPLANTATION PERIOD. INCREASING AND / OR PERMANENT LOAD APPLICATION WILL INEVITABLY LEAD TO FATIGUE OF MATERIAL. EVALUATION REVEALED EVIDENCE THAT THE EVENT IS NOT LINKED TO A DEFICIENCY OF THE DEVICE BUT IS RATHER RELATED TO HIGH LOAD APPLICATION BY THE PATIENT CONTRIBUTED BY INTRA OPERATIVE DAMAGE OF THE PROXIMAL DRILL HOLE. THE REVIEW OF THE RISK ASSESSMENT FOR THE FAILURE MODE AND LEVEL 1 ROOT CAUSE INDICATED THE ISSUE WAS WITHIN TOLERANCE, THEREFORE, NO CORRECTIVE ACTIONS ARE DEEMED NECESSARY AT THIS TIME.
GAMMA NAIL BROKE IN PATIENT.
GAMMA NAIL BROKE IN PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 245052 | LONG NAIL KIT R1.5, TI, RIGHT GAMMA3 Ø10X360MM X 125° | IMPLANT | HSB | STRYKER OSTEOSYNTHESIS-KIEL | K900066 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |