FDA Adverse Event Injury Summary report: N

LONG NAIL KIT R1.5, TI, RIGHT GAMMA3 Ø10X360MM X 125°

MDR report key: 3141327 · Received June 3, 2013

Report

Report Number
0009610622-2013-00308
Event Type
Injury
Date Received
June 3, 2013
Date of Event
May 1, 2013
Report Date
May 13, 2013
Manufacturer
STRYKER OSTEOSYNTHESIS-KIEL
Product Code
HSB
PMA / PMN Number
K034002
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ONCE THE INVESTIGATION HAS BEEN COMPLETED ANY ADDITIONAL INFORMATION WILL BE REPORTED IN A SUPPLEMENTAL REPORT.

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: THE RETURNED DEVICE MATCHED THE REPORT, AND THE INCIDENT WAS CONFIRMED. INVESTIGATION REVEALED NO DISCREPANCIES IN MATERIAL OF MANUFACTURING. ACCORDING TO THE DAMAGE AND THE EVIDENCES OF THE BREAKAGE SURFACES THE NAIL BROKE IN A FATIGUE FRACTURE DUE TO OVERLOAD, CONTRIBUTED BY INTRA OPERATIVELY DAMAGE OF THE NAIL CAUSED BY A DEVIATED LAG SCREW STEP DRILL. DUE TO A NOTCHING EFFECT, THE MISS DRILLING MOST LIKELY HAD CREATED THE STARTING POINT OF THE IMPLANT BREAKAGE AT THE LATERAL EDGE OF THE ANTERIOR BRIDGE. WE DID NOT RECEIVE A REASONABLE STATEMENT REGARDING POSTOPERATIVE MOBILIZATION INSTRUCTIONS / WEIGHT BEARING. NEVERTHELESS, AS THE PATIENT WAS REVISED IT CAN BE ASSUMED THAT BONE CONSOLIDATION WAS INSUFFICIENT. CONSEQUENTLY, THE NAIL WAS NOT RELIEVED BY BONE CONSOLIDATION AND HAD TO ABSORB THE COMPLETE LOAD APPLICATION DURING THE IMPLANTATION PERIOD. INCREASING AND / OR PERMANENT LOAD APPLICATION WILL INEVITABLY LEAD TO FATIGUE OF MATERIAL. EVALUATION REVEALED EVIDENCE THAT THE EVENT IS NOT LINKED TO A DEFICIENCY OF THE DEVICE BUT IS RATHER RELATED TO HIGH LOAD APPLICATION BY THE PATIENT CONTRIBUTED BY INTRA OPERATIVE DAMAGE OF THE PROXIMAL DRILL HOLE. THE REVIEW OF THE RISK ASSESSMENT FOR THE FAILURE MODE AND LEVEL 1 ROOT CAUSE INDICATED THE ISSUE WAS WITHIN TOLERANCE, THEREFORE, NO CORRECTIVE ACTIONS ARE DEEMED NECESSARY AT THIS TIME.

Description of Event or Problem · 1

GAMMA NAIL BROKE IN PATIENT.

Description of Event or Problem · 1

GAMMA NAIL BROKE IN PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
245052 LONG NAIL KIT R1.5, TI, RIGHT GAMMA3 Ø10X360MM X 125° IMPLANT HSB STRYKER OSTEOSYNTHESIS-KIEL K900066

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention