FDA Adverse Event Malfunction Summary report: N

ACCU-CHEK SPIRIT COMBO

MDR report key: 3141297 · Received June 3, 2013

Report

Report Number
2183996-2013-00997
Event Type
Malfunction
Date Received
June 3, 2013
Date of Event
May 26, 2013
Report Date
June 25, 2013
Manufacturer
ROCHE DIABETES CARE AG
Product Code
LZG
PMA / PMN Number
NA
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THE SOFTCOMPONENTS OF THE HOUSING ARE WORN DOWN HEAVILY AND RESTRICTED HANDLING OF THE PUMP IS A POSSIBLE IMPLICATION. THEREFORE MOISTURE ENTERED THE INSULIN PUMP AND DESTROY THE FUNCTIONALITY OF THE BUTTONS AND THE PUMP ELECTRONICS. THE UP BUTTON IS PERMANENT ACTIVE DUE TO AN EXTERNAL MECHANIC DAMAGE.

Additional Manufacturer Narrative · 1

THE INCIDENT OCCURRED OUTSIDE THE UNITED STATES. INFORMATION CONTAINED WITHIN THIS REPORT IS ALL THAT IS AVAILABLE AT THIS TIME. IF FURTHER INFORMATION IS OBTAINED, IT WILL BE PROVIDED IN THE SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

ON (B)(6) 2013, IT WAS REPORTED THAT THE PATIENT'S INFUSION DEVICE DISPLAYED E8 (POWER INTERRUPT), BUT SHE WAS UNABLE TO CLEAR THE ERROR BECAUSE THE CHECK BUTTON WAS NOT FUNCTIONING. NO PHYSIOLOGICAL EFFECTS WERE REPORTED. THE PATIENT DID NOT REQUIRE MEDICAL ASSISTANCE FROM A HEALTHCARE PROFESSIONAL OR SECOND PARTY TO ADDRESS THE ISSUE. THE INFUSION DEVICE WAS REQUESTED TO BE RETURNED FOR PRODUCT EVALUATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
245022 ACCU-CHEK SPIRIT COMBO INSULIN INFUSION DEVICE LZG ROCHE DIABETES CARE AG 00700006863 NA

Patients

Seq Age Sex Outcome Treatment
1 023 YR