FDA Adverse Event Malfunction Summary report: N

LOCKING SCREW VARIAX FULL THREAD 2.7MM / L16MM

MDR report key: 3141292 · Received June 3, 2013

Report

Report Number
0008031020-2013-00169
Event Type
Malfunction
Date Received
June 3, 2013
Date of Event
May 7, 2013
Report Date
May 7, 2013
Manufacturer
STRYKER OSTEOSYNTHESIS-SELZACH
Product Code
HRS
PMA / PMN Number
K101056
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ONCE THE INVESTIGATION HAS BEEN COMPLETED ANY ADDITIONAL INFORMATION WILL BE REPORTED IN A SUPPLEMENTAL REPORT.

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: BASED ON INVESTIGATION, THE ROOT CAUSE OF THE DETERMINED EVENT WAS ATTRIBUTED TO A USER RELATED ISSUE. THE THREAD ON THE SCREW HEAD WAS DAMAGED BY ANOTHER ITEM. IF THE THREAD ON THE HEAD IS DAMAGED THE LOCKING MECHANISM DOESN'T WORK PROPERLY. PLEASE FOLLOW EXACTLY THE OP-TECH TO LOCK THE SCREW IN THE PLATE AND DO NOT RECOMMEND ITEMS WITH THE SCREW. A REVIEW OF THE DEVICE HISTORY FOR THE REPORTED LOT DID NOT INDICATE ANY ABNORMALITIES. NO CORRECTIVE ACTIONS ARE REQUIRED AT THIS TIME. INDICATIONS FOR ANY MATERIAL, MANUFACTURING OR DESIGN RELATED PROBLEMS WERE NOT DETERMINED IN THE INVESTIGATION.

Description of Event or Problem · 1

DURING VARIAX ELBOW SURGERY, ALTHOUGH THE SURGEON INSERTED THE LOCKING SCREW IN PLATE SCREW HOLE, THE LOCKING SCREW WAS NOT ABLE TO BE LOCKED TO PLAT E(THE SURGEON TRIED ABOUT 5 TIMES). THEREFORE, THE SURGEON CHANGED THE LOCKING SCREW TO ANOTHER OF THE SAME SIZE AND COMPLETED THE OPERATION.

Description of Event or Problem · 1

DURING VARIAX ELBOW SURGERY, ALTHOUGH THE SURGEON INSERTED THE LOCKING SCREW IN PLATE SCREW HOLE, THE LOCKING SCREW WAS NOT ABLE TO BE LOCKED TO PLATE(THE SURGEON TRIED ABOUT 5 TIMES). THEREFORE THE SURGEON CHANGED THE LOCKING SCREW TO ANOTHER OF THE SAME SIZE AND COMPLETED THE OPERATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
244397 LOCKING SCREW VARIAX FULL THREAD 2.7MM / L16MM IMPLANT HRS STRYKER OSTEOSYNTHESIS-SELZACH Z27803

Patients

Seq Age Sex Outcome Treatment
1 51 YR Other