LH 500 HEMATOLOGY ANALYZER
Report
- Report Number
- 1061932-2013-00940
- Event Type
- Malfunction
- Date Received
- June 3, 2013
- Date of Event
- May 3, 2013
- Report Date
- May 3, 2013
- Manufacturer
- BECKMAN COULTER
- Product Code
- GKZ
- PMA / PMN Number
- K032000
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
A BECKMAN COULTER FSE (FIELD SERVICE ENGINEER) WAS DISPATCHED AND FOUND A HOLE IN THE TUBING THROUGH PINCH VALVE PV35. THE FSE REPLACED THE TUBING AND NO FURTHER LEAKS WERE OBSERVED. THE FSE DISCOVERED THAT THE LEAK HAD DAMAGED THE PELTIER MODULE AND CAUSED THE INSTRUMENT TO GENERATED TEMPERATURE ERRORS. THE FSE PROCEEDED TO REPLACE THE PELTIER MODULE TO RESOLVE THE TEMPERATURE ERRORS AND REPAIRS WERE VERIFIED PER ESTABLISHED PROCEDURES. FAILURE MODE OF THE LEAK IS ATTRIBUTED TO A HOLE IN THE TUBING THROUGH PINCH VALVE PV35 AND FAILURE MODE OF THE TEMPERATURE ERRORS IS ATTRIBUTED TO A DAMAGED PELTIER MODULE. THE INSTRUMENT OPERATED AS INTENDED BY GENERATING TEMPERATURE ERRORS TO ALERT THE OPERATOR TO AN INSTRUMENT PROBLEM. (B)(4).
THE CUSTOMER REPORTED A LEAK OF APPROXIMATELY 5 ML INSIDE THE LH 500 HEMATOLOGY ANALYZER. THE CUSTOMER INDICATED THAT THE LEAK WAS CONTAINED WITHIN THE INSTRUMENT AND THE INSTRUMENT GENERATED TEMPERATURE ERRORS WHEN THE LEAK WAS NOTICED. THE CUSTOMER WAS WEARING PERSONAL PROTECTIVE EQUIPMENT CONSISTING OF GLOVES AND A LABORATORY COAT AT THE TIME OF THE EVENT AND NO INJURY OR EXPOSURE WAS REPORTED. NO DISCREPANT PATIENT RESULTS WERE GENERATED AND THERE WAS NO CHANGE OR AFFECT TO PATIENT TREATMENT IN CONNECTION WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 243025 | LH 500 HEMATOLOGY ANALYZER | COUNTER, DIFFERENTIAL CELL | GKZ | BECKMAN COULTER | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |