FDA Adverse Event Injury Summary report: N

CHARGING SYSTEM

MDR report key: 3141090 · Received May 24, 2013

Report

Report Number
1627487-2013-03700
Event Type
Injury
Date Received
May 24, 2013
Date of Event
May 2, 2013
Report Date
May 2, 2013
Manufacturer
ST. JUDE MEDICAL - NEUROMODULATION
Product Code
LGW
PMA / PMN Number
P010032
Removal / Correction Number
1627487-07262012-001-C
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THIS CHARGER MODEL WAS ASSOCIATED WITH A FIELD CORRECTION. MFR'S EVAL: CORRECTIVE AND PREVENTIVE ACTION (CAPA) INVESTIGATION PERFORMED. RESULT: POCKET HEATING WAS CONFIRMED. THE INVESTIGATION FOR CAPA (B)(4) ASSOCIATED WITH HEATING WHILE CHARGING (POCKET HEATING) CONCLUDED THAT THE CHARGER WAS CAPABLE OF TRANSFERRING ENERGY TO THE IPG AT A RATE THAT WOULD CAUSE HEATING OF THE IPG AND/OR CHARGING WAND OF SUFFICIENT ELEVATED TEMP TO CAUSE PAIN AND BURNS. THE HEATING WHILE CHARGING WAS DETERMINED TO BE EXACERBATED BY OFF-AXIS CHARGING OF SHALLOW IMPLANTED IPGS AND THAT ALL CHARGERS WERE CAPABLE OF ELEVATED HEATING. SJM HAS LIMITED INFO RELATED TO THE PT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Description of Event or Problem · 1

DEVICE 2 OF 2. REF MFR REPORT: 1627487-2013-03699. THE PT REPORTED EXPERIENCING WARMING AT HER SCS IPG POCKET SITE WHILE CHARGING. SUBSEQUENTLY, A REPLACEMENT CHARGING SYSTEM (LOW ENERGY) WAS SENT TO THE PT. THE PT ALSO REPORTED DISCOMFORT AT THE POCKET SITE REGARDLESS OF STIMULATION (REF MFR REPORTS: 1627487-2013-13737 AND 1627487-2013-13738). F/U IDENTIFIED THE POCKET HEATING ISSUE RESOLVED WITH THE REPLACEMENT CHARGING SYSTEM. ON 08/01/2012 ST JUDE MEDICAL, NEUROMODULATION DIV., SENT FIELD ACTION LETTERS TO PTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS ARE EXPECTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
230644 CHARGING SYSTEM SCS CHARGING SYSTEM LGW ST. JUDE MEDICAL - NEUROMODULATION 3721 3097343

Patients

Seq Age Sex Outcome Treatment
1 56 YR Required Intervention IMPLANT DATE:| SCS ANCHOR: MODEL 1194 (2)| IMPLANT DATE:| SCS LEAD: MODEL 3186 (2)