FDA Adverse Event Malfunction Summary report: N

HOMECHOICE

MDR report key: 3140962 · Received June 3, 2013

Report

Report Number
1416980-2013-13987
Event Type
Malfunction
Date Received
June 3, 2013
Date of Event
April 22, 2013
Report Date
May 10, 2013
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION: THE DEVICE WAS RETURNED TO BAXTER, AND THE EVALUATION IS COMPLETE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT OPENED DURING INVESTIGATION OF (B)(4). THE REPORTED DIFFICULTY OF AN IIPV EVENT WAS CONFIRMED THROUGH AN EVENT HISTORY LOG REVIEW. THE CAUSE WAS FALSE EMPTY DETECT AND USE ERROR, CLINICIAN INAPPROPRIATELY SET MINIMUM DRAIN VOLUME PERCENT SETTING TOO LOW. THE HOMECHOICE APD SYSTEMS TRAINER'S GUIDE PROVIDES A WARNING "IF YOU SET THE MINIMUM DRAIN VOLUME % TOO LOW, AN INCOMPLETE DRAIN COULD RESULT FOR THE PATIENT. THIS COULD RESULT IN AN INCREASED INTRAPERITONEAL VOLUME (IIPV) SITUATION." IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A FOLLOW-UP WILL BE SENT.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE MACHINE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT(S) WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2013 AT 09:07:13. DURING NIGHT DRAIN CYCLE FIFTEEN, THE PATIENT'S ULTRAFILTRATION READING WAS 1284ML, INDICATING THE HOME PATIENT (HP) DRAINED 1284ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 2000ML. THIS INFORMATION MEETS IIPV CRITERIA. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
245033 HOMECHOICE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1