FDA Adverse Event Malfunction Summary report: N

INFUSOR

MDR report key: 3140804 · Received May 31, 2013

Report

Report Number
1416980-2013-13967
Event Type
Malfunction
Date Received
May 31, 2013
Date of Event
May 7, 2013
Report Date
May 7, 2013
Manufacturer
BAXTER HEALTHCARE - IRVINE
Product Code
MEB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE IS AVAILABLE FOR EVALUATION; HOWEVER, THE DEVICE HAS NOT YET BEEN RECEIVED BY BAXTER. A SUPPLEMENTAL REPORT WILL BE FILED IF ANY ADDITIONAL RELEVANT INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). THE RESERVOIR WAS ALSO MICROSCOPICALLY EXAMINED FOR ANY ABNORMALITIES THAT MAY HAVE POTENTIALLY CONTRIBUTED TO THE RUPTURE. AN ABRASION WAS DETECTED ON THE INTERIOR SURFACE OF THE RESERVOIR NEAR THE RUPTURE LINE. THE EXACT CAUSE OF THE REPORTED CONDITION WAS NOT DETERMINED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). A REVIEW OF ALL BATCH RECORD DOCUMENTS WAS PERFORMED WITH NO ISSUES NOTED DURING THE MANUFACTURING PROCESS. THERE WERE NO DEVIATIONS FROM STANDARD PROCEDURE AND NO EXCEPTIONS RELATED TO THE REPORTED CONDITION WERE NOTED. THE SAMPLE WAS RECEIVED FOR EVALUATION. DURING VISUAL INSPECTION, A RUPTURED BLADDER IN A NON-FOOTING POSITION WAS IDENTIFIED. THE REPORTED CONDITION WAS CONFIRMED. UPON COMPLETION OF BAXTER'S INVESTIGATION, THEN A FOLLOW-UP MDR WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE BLADDER OF AN INFUSOR SV2.5 DEVICE RUPTURED DURING FILLING. THE DRUG WAS FILLED MANUALLY BY A SYRINGE. THERE WAS NO PATIENT INVOLVEMENT REPORTED. ADDITIONAL INFORMATION WAS REQUESTED AND IS NOT AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
242322 INFUSOR PUMP, INFUSION, ELASTOMERIC MEB BAXTER HEALTHCARE - IRVINE 13A009

Patients

Seq Age Sex Outcome Treatment
1