FDA Adverse Event Malfunction Summary report: N

ARCHITECT CA 19-9XR

MDR report key: 3140683 · Received May 31, 2013

Report

Report Number
1415939-2013-00255
Event Type
Malfunction
Date Received
May 31, 2013
Date of Event
May 10, 2013
Report Date
May 10, 2013
Manufacturer
ABBOTT LABORATORIES
Product Code
NIG
PMA / PMN Number
K052000
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SI
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

AN EVALUATION IS IN PROCESS. A FOLLOW UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. (B)(4).

Additional Manufacturer Narrative · 1

FURTHER INVESTIGATION OF THE CUSTOMER ISSUE INCLUDED A REVIEW OF THE COMPLAINT INFORMATION, A SEARCH FOR SIMILAR COMPLAINTS, A REVIEW OF LABELING AND ACCURACY TESTING. NO ADVERSE TREND WAS IDENTIFIED FOR THE CUSTOMER'S ISSUE. LABELING WAS REVIEWED AND FOUND TO BE ADEQUATE. ACCURACY TESTING OF PANELS OF KNOWN ANALYTE CONCENTRATIONS MET ALL SPECIFICATIONS AND ACCEPTANCE CRITERIA. COMPLAINT INFORMATION REASONABLY SUGGESTS THE ASSAY IS PERFORMING AS INTENDED, AND NO MALFUNCTION OF THE DEVICE OCCURRED. BASED ON THE AVAILABLE INFORMATION, NO PRODUCT DEFICIENCY WAS IDENTIFIED.

Description of Event or Problem · 1

THE CUSTOMER OBSERVED A FALSELY ELEVATED CA19-9 RESULT FOR ONE PATIENT ON THE ARCHITECT I2000SR ANALYZER. THE FOLLOWING DATA WAS PROVIDED FROM (B)(6) 2013: SID (B)(6) AT 8:16 FIRST TIME MEASURED 48.93 U/ML, AT 8:59 SECOND TIME MEASURED 10.10, AT 9:35 THIRD TIME MEASURED 11.93 (ALL 3 TIMES SPECIMEN ASPIRATED FROM THE SAME PRIMARY TUBE), NORMAL RANGE IS 0-37 U/ML, THEREFORE, THE FIRST RESULT POSITIVE, SECOND AND THIRD RESULT NEGATIVE. ABBOTT CONTROLS WERE TESTED LOW, MED AND HIGH AND ALL IN RANGE. ABBOTT TECHNICAL SUPPORT VISITED THE CUSTOMER AND CHECKED THE INSTRUMENT PERFORMANCE, VALVE PRESSURE, CLEANED THE PRETRIGGER/TRIGGER MANIFOLD AND THERE WAS NO INSTRUMENT LEAKAGE FOUND.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
241591 ARCHITECT CA 19-9XR NIG ABBOTT LABORATORIES 21121M500

Patients

Seq Age Sex Outcome Treatment
1 LN 03M74-01 SN (B)(4)| ARCHITECT I2000SR ANALYZER