VISIAN ICL (IMPLANTABLE COLLAMER LENS)
Report
- Report Number
- 2023826-2013-00438
- Event Type
- Injury
- Date Received
- May 31, 2013
- Report Date
- May 6, 2013
- Manufacturer
- STAAR SURGICAL COMPANY
- Product Code
- MTA
- PMA / PMN Number
- P030016
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). DEVICE EVALUATED BY MANUFACTURER? NO - 81 (OTHER): LENS WAS DISCARDED. METHOD: WORK ORDER SEARCH. RESULTS: A LENS WORK ORDER SEARCH WAS PERFORMED AND NO SIMILAR COMPLAINTS WERE FOUND WITHIN THE WORK ORDER. CONCLUSIONS: (NO CONCLUSION CAN BE DRAWN): BASED ON THE COMPLAINT HISTORY AND WORK ORDER SEARCH, A SPECIFIC ROOT CAUSE OF THE EVENT COULD NOT BE DETERMINED. (B)(4). LENS WAS DISCARDED.
RESULTS: MEDICAL REVIEW - REPORTEDLY ICL WAS EXPLANTED/EXCHANGED FOR A SHORTER LENS TWO WEEKS POSTOPERATIVELY TO ADDRESS ELEVATED IOP( NO VALUE PROVIDED).AFTER THE INTERVENTION THE IOP WAS 14MMHG. ACCORDING TO THE REPORT THE EXACT CAUSE OF THE EVENT WAS UNKNOWN. GIVEN THE INFORMATION THAT SHORTER LENS RESOLVED THE ISSUE THE MOST LIKELY REASON WAS PATIENT RELATED FACTOR(ANATOMY),COMBINED WITH PROCEDURE RELATED FACTORS ((B)(4)). CONCLUSIONS: BASED ON THE COMPLAINT HISTORY, WORK ORDER SEARCH AND THE MEDICAL REVIEW, A SPECIFIC ROOT CAUSE OF THE EVENT COULD NOT BE DETERMINED. (B)(4).
THE REPORTER STATED THE SURGEON IMPLANTED A 13.2MM MICL 13.2 IMPLANTABLE COLLAMER LENS IN THE PATIENT'S LEFT EYE (OS) ON (B)(6) 2013. THE LENS WAS EXPLANTED ON (B)(6) 2013, DUE TO ELEVATED IOP. THE LENS WAS REMOVED WITH NO PATIENT INJURY AND WAS DISCARDED. THE LENS WAS EXCHANGED WITH A SHORTER LENS. THE IOP AFTER EXPLANT AND EXCHANGE WAS 14MM AND THE PATIENT IS VERY HAPPY WITH THE IMPLANT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 241509 | VISIAN ICL (IMPLANTABLE COLLAMER LENS) | INTRAOCULAR LENS | MTA | STAAR SURGICAL COMPANY | MICL 13.2 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 26 YR | Required Intervention | CARTRIDGE MODEL SFC-45 FP - LOT NUMBER UNK.| CARTRIDGE MODEL-SFC-45, LOT#-UNK.| FOAM TIP PLUNGER MODEL FTP - LOT NUMBER UNK.| FOAM TIP PLUNGER MODEL-FTP, LOT#-UNK.| INJECTOR MODEL MSI-PF - LOT NUMBER UNK.| INJECTOR MODEL-MSI-PF, LOT#-UNK.| INJECTOR MODEL MSI-PF - LOT NUMBER UNK| CARTRIDGE MODEL SFC-45 FP - LOT NUMBER UNK| FOAM TIP PLUNGER MODEL FTP - LOT NUMBER UNK |