ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-07451
- Event Type
- Malfunction
- Date Received
- May 31, 2013
- Report Date
- May 24, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
FOLLOW-UP #1 DATE OF SUBMISSION 08/20/2013-DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 07/25/2013 WITH THE FOLLOWING FINDINGS: EVALUATION REVEALED ONE LOW BATTERY ALARM IN BLACK BOX. THE BATTERY COMPARTMENT WAS INTACT. NO EVIDENCE OF MOISTURE CONTAMINATION FOUND INSIDE BATTERY COMPARTMENT AND THE BATTERY CAP IS ABLE TO FULLY TIGHTEN. THE CURRENT DRAWS WERE WITHIN SPECIFICATIONS. THE DISPLAY WAS FOUND TO BE FADED, DISCOLORED AND DIFFICULT TO READ. INVESTIGATORS WERE UNABLE TO DUPLICATE OR VERIFY EXCESSIVE BATTERY USAGE.
ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (BATTERY LIFE) ISSUE. THE REPORTER STATED THAT THEY CHANGED BATTERY YESTERDAY AND PUMP JUST DIED WITHOUT LOW BATTERY ALARM. ALARM HISTORY INDICATED BATTERY CHANGES OCCURRING FREQUENTLY. THE REPORTER DENIED MOISTURE OR TRAUMA AND NO CRACKS IN BATTERY COMPARTMENT. THE BATTERY CAP WAS INTACT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 240711 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 3 YR |