FDA Adverse Event Malfunction Summary report: N

UV FLASH TRANSFER SETS

MDR report key: 3140471 · Received May 31, 2013

Report

Report Number
1416980-2013-13939
Event Type
Malfunction
Date Received
May 31, 2013
Date of Event
May 20, 2013
Report Date
May 20, 2013
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
KDJ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). AN EVALUATION OF THE ACTUAL SAMPLE WAS CONDUCTED. VISUAL INSPECTION, LEAK TESTING AND CLEAR PASSAGE TESTING WAS PERFORMED WITH NO ISSUES NOTED. FUNCTIONALLY HAND TIGHTENED A LAB TITANIUM ADAPTER TO SET WITH DIFFICULTY NOTED. SAMPLE WAS CONFIRMED FOR THE REPORTED PROBLEM. DIMENSIONAL INSPECTION OF THE RETURNED TRANSFER SET SAMPLE IDENTIFIED THAT THE INSIDE DIAMETER OF THE CATHETER ADAPTER SHROUD WAS BELOW NOMINAL. IMPROVEMENTS WERE MADE TO THE MOLD AND MOLDING DEPARTMENT PROCEDURES. A FOLLOW-UP REPORT WILL BE FILED IF ANY ADDITIONAL RELEVANT INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE IS REPORTED TO BE AVAILABLE BUT HAS NOT BEEN RECEIVED AT THIS TIME. A REVIEW OF ALL BATCH RECORD DOCUMENTS WAS PERFORMED WITH NO ISSUES NOTED DURING THE MANUFACTURING PROCESS. THERE WERE NO DEVIATIONS FROM STANDARD PROCEDURE AND NO EXCEPTIONS RELATED TO THE REPORTED CONDITION WERE NOTED. SHOULD THE DEVICE BE RECEIVED OR ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THIS IS THE SAME PATIENT AS (B)(4).

Description of Event or Problem · 1

THIS IS REPORT 2 OF 4 FOR THIS EVENT. IT WAS REPORTED THAT THE PATIENT CONNECTOR ON A TRANSFER SET WOULD NOT CONNECT WELL WITH THE TITANIUM ADAPTER WHEN THE TRANSFER SET WAS CHANGED. THERE WAS PATIENT INVOLVEMENT BUT NO PATIENT INJURY OR MEDICAL INTERVENTION WAS INDICATED IN ASSOCIATION WITH THIS REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
240722 UV FLASH TRANSFER SETS SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE KDJ BAXTER HEALTHCARE - MOUNTAIN HOME H13A02017

Patients

Seq Age Sex Outcome Treatment
1 TITANIUM ADAPTER