PROGRAMMABLE VALVE IN-LINE W/ SIPHONGUARD
Report
- Report Number
- 1226348-2013-17931
- Event Type
- Injury
- Date Received
- May 31, 2013
- Date of Event
- May 13, 2013
- Manufacturer
- CODMAN & SHURTLEFF, INC.
- Product Code
- JXG
- PMA / PMN Number
- PK992173
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
UPON COMPLETION OF THE INVESTIGATION IT WAS NOTED THAT DURING THE VISUAL EXAMINATION OF THE DEVICE DEEP MARKS WERE NOTED IN THE SILICONE HOUSING AROUND THE END OF THE SIPHON GUARD, THE MARKS WERE NOTED ON OPPOSITE SIDES, WHICH MIGHT INDICATE THAT THE SIPHON GUARD WAS CLAMPED TOO TIGHT, THIS HOWEVER COULD NOT BE DETERMINED. IN ADDITION, A HOLE AND TEAR WAS ALSO FOUND IN THE NEEDLE CHAMBER. IT IS NOT POSSIBLE TO DETERMINE HOW THIS HAS HAPPENED. FURTHER EXAMINATION ALSO REVEALED THAT SINCE BIOLOGICAL DEBRIS WAS FOUND IN THE VALVE CASING IT WAS NOT POSSIBLE TO SEE THE CAM POSITION. THE VALVE COULD NOT BE IRRIGATED DUE TO THE DAMAGE OF THE SILICONE HOUSING. THE VALVE COULD NOT BE PROGRAM TESTED AS THE VALVE CAM MECHANISM WAS NOT VISIBLE DUE TO BIOLOGICAL DEBRIS. DUE TO THE DAMAGE IN THE SILICONE HOUSING, IT WAS NOT POSSIBLE TO PRESSURE TEST THE VALVE. A REVIEW OF THE DEVICE HISTORY RECORDS CONFIRMED THAT THIS DEVICE CONFORMED TO THE REQUIRED SPECIFICATIONS PRIOR TO DISTRIBUTION. BASED ON THE RESULTS OF THIS INVESTIGATION NO FURTHER ACTION IS REQUIRED. TRENDS WILL BE MONITORED FOR THIS AND SIMILAR COMPLAINTS. AT THE PRESENT TIME, THIS COMPLAINT IS CLOSED.
UPON COMPLETION OF THE INVESTIGATION, A FOLLOW UP REPORT WILL BE FILED.
THE AFFILIATE REPORTED THAT THE DEVICE WAS IMPLANTED AT 80MM H2O. THE PATIENT¿S VENTRICLES BECAME LARGE DUE TO A HEAD INJURY. THE PRESSURE WAS CHANGED FROM 80MM H2O TO 60MM H2O BUT THE PATIENT¿S CONDITION DID NOT CHANGE. THE PRESSURE WAS THEN CHANGED TO 40MM H2O AND THERE WAS STILL NO IMPROVEMENT. LASTLY, THE PRESSURE WAS CHANGED TO 30MMH2O WITH NO IMPROVEMENT IN THE PATIENT¿S CONDITION. AS A RESULT, THE DEVICE WAS REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 242331 | PROGRAMMABLE VALVE IN-LINE W/ SIPHONGUARD | SHUNT, CENTRAL NERVOUS SYSTEM & COMPS | JXG | CODMAN & SHURTLEFF, INC. | CMHB4K |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 24 YR | Required Intervention |