ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2013-07449
- Event Type
- Malfunction
- Date Received
- May 31, 2013
- Date of Event
- May 6, 2013
- Report Date
- May 7, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME. (B)(6).
FOLLOW-UP # 1 DATE OF SUBMISSION 07/26/2013-DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: THE DISPLAY SCREEN WAS OBSERVED TO BE DIM, DISCOLORED, AND DIFFICULT TO READ. A TEST SCREEN WAS INSTALLED AND DISPLAYED PROPERLY. UNRELATED TO THE COMPLAINT, THE AUDIO BOLUS BUTTON WAS FOUND TO BE UNRESPONSIVE. THE COVER WAS REMOVED AND THE CONTACT WAS FOUND TO BE MISALIGNED.
ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (DIM/FADING/COLOR SPECTRUM) ISSUE. IT WAS REPORTED THAT THE DISPLAY SCREEN WAS DIM, FADED AND DISCOLORED. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.