FDA Adverse Event Malfunction Summary report: N

GENESYS HTA PROCERVA®

MDR report key: 3140458 · Received May 31, 2013

Report

Report Number
3005099803-2013-04594
Event Type
Malfunction
Date Received
May 31, 2013
Date of Event
May 10, 2013
Report Date
May 10, 2013
Manufacturer
BOSTON SCIENTIFIC - MARLBOROUGH
Product Code
MNB
PMA / PMN Number
P000040
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NE, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ACCORDING TO THE COMPLAINANT, THE DEVICE WAS DISPOSED OF AND IS NOT AVAILABLE FOR EVALUATION. AT THIS TIME, WE ARE UNABLE TO DETERMINE THE RELATIONSHIP BETWEEN THE DEVICE AND THE CAUSE FOR THIS EVENT. IF THERE IS ANY FURTHER RELEVANT INFORMATION RECEIVED, A SUPPLEMENTAL MEDWATCH REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A GENESYS HYDROTHERMABLATION PROCERVA PROCEDURE SET WAS USED IN A HYDROTHERMABLATION (HTA) PROCEDURE PERFORMED ON (B)(6) 2013. ACCORDING TO THE COMPLAINANT, THE PATIENT WAS PREOPERATIVELY PRESCRIBED CYTOTEC AND WAS REPORTED TO HAVE A STENOTIC CERVIX THAT WAS BARELY DILATED FOR THE PROCEDURE. DURING THE PROCEDURE THEY HAD A 10 CC FLUID LOSS AT 3 MINUTES INTO ABLATION AND AT EIGHT MINUTES INTO ABLATION, A FLUID LOSS A 10CC WAS AGAIN REPORTED WITH NO VISIBLE FLUID LOSS FOR EITHER AT THE CERVICAL OS. PHYSICIAN APPLIED A SECOND TENACULUM AT THE CERVIX AT THE 3 O'CLOCK POSITION. THE PHYSICIAN RECEIVED A 3RD 10CC FLUID LOSS ALARM AT THE SAME TIME PATIENT COMMENTED THAT SHE FELT A HOT SENSATION IN HER VAGINA. COLD SALINE WAS IMMEDIATELY APPLIED AND THE MACHINE WENT INTO COOL DOWN. COOL SALINE WAS THEN POURED INTO THE VAGINA. UPON HYSTEROSCOPY NO EVIDENCE OF BURN WAS NOTED AND NO TREATMENT WAS PROVIDED FOR THE ISSUE. THERE WERE NO REPORTED PATENT COMPLICATIONS AND THE PATIENT'S CONDITION HAS BEEN DESCRIBED AS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
241267 GENESYS HTA PROCERVA® DEVICE, THERMAL ABLATION, ENDOMETRIAL MNB BOSTON SCIENTIFIC - MARLBOROUGH M006580210 0000044396

Patients

Seq Age Sex Outcome Treatment
1 50 YR