FDA Adverse Event Malfunction Summary report: N

3.5MM 90-S SERFAS ENERGY SUCTION PROBE

MDR report key: 3140412 · Received May 31, 2013

Report

Report Number
0002936485-2013-00239
Event Type
Malfunction
Date Received
May 31, 2013
Date of Event
April 30, 2013
Report Date
May 3, 2013
Manufacturer
STRYKER ENDOSCOPY-SAN JOSE
Product Code
GEI
PMA / PMN Number
K071859
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCT WAS RETURNED FOR INVESTIGATION. THE REPORTED FAILURE MODE WAS CONFIRMED. SCRATCH WEAR MARKS WERE OBSERVED ON ALL OF THE CERAMIC AND LUMEN. NO VISUAL WEAR MARKS WERE OBSERVED ON THE PROBE¿S INSULATION. REVIEW OF THE DEVICE HISTORY RECORD OF THE REPORTED LOT NUMBER WAS REVIEWED. THE REVIEW DISCLOSED NO DISCREPANCIES THAT COULD CONTRIBUTE TO THIS CONDITION. BASED ON THE RETURNED UNIT'S EVALUATION, THE MOST PROBABLE ROOT CAUSE COULD BE DESIGN (STRENGTH) COMBINED WITH USER RELATED (MISUSE). THE WEAR MARKS ARE INDICATIVE OF FRICTION OR SCRAPPING WITH ANOTHER POINTY HARD OR RIGID OBJECT OR DEVICE DURING SURGERY (E.G. CUTTER OR SHAVER, HARD TISSUE OR BONE), POSSIBLY WHEN INSERTING OR REMOVING THE PROBE INTO AN OBSTRUCTED PASSAGEWAY. IT IS POSSIBLE THAT THE PROBE WAS USED AS A TOOL FOR THE MECHANICAL DISPLACEMENT OF TISSUE OR BONE, OR AS A PRYING OR SCRUBBING TOOL, WHICH MAY RESULT IN DAMAGE TO THE TIP. IN SUM, THE PRODUCT WAS RETURNED FOR INVESTIGATION AND THE REPORTED FAILURE MODE WAS CONFIRMED. THE FAILURE MODE WILL BE MONITORED FOR FUTURE REOCCURRENCE.

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION WILL BE PROVIDED ONCE THE INVESTIGATION HAS BEEN COMPLETED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PROCEDURE, THE TIP OF THE DEVICE BROKE OFF INSIDE THE JOINT OF THE PATIENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PROCEDURE, THE TIP OF THE DEVICE BROKE OFF INSIDE THE JOINT OF THE PATIENT.