COAGUCHEK ® XS SYSTEM
MDR report key: 3140367
·
Received May 31, 2013
Report
- Report Number
- 1823260-2013-03288
- Event Type
- Malfunction
- Date Received
- May 31, 2013
- Date of Event
- April 1, 2013
- Report Date
- June 27, 2013
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- GJS
- PMA / PMN Number
- K062925
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
IT WAS UNKNOWN IF THE INITIAL REPORTER SENT REPORT TO THE FDA.
Description of Event or Problem · 1
CALLER REPORTS THAT DURING A CORRELATION STUDY THE FOLLOWING COAGUCHEK XS/LABORATORY RESULTS WERE OBTAINED: 4.5 INR/2.8 INR, 6.1 INR/4.2 INR, 4.2 INR/3.1 INR, 4.5 INR/3.4 INR, 4.8 INR/3.5 INR. NO ACTIONS TAKEN BASED ON DEVICE RESULTS. NO ADVERSE EVENT REPORTED. REQUESTED RETURN OF SUSPECT SYSTEM AND REPLACEMENT WAS SENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 242151 | COAGUCHEK ® XS SYSTEM | PROTHROMBIN TIME TEST STRIPS | GJS | ROCHE DIAGNOSTICS | NA | 21637312 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |