FDA Adverse Event Malfunction Summary report: N

COAGUCHEK ® XS SYSTEM

MDR report key: 3140367 · Received May 31, 2013

Report

Report Number
1823260-2013-03288
Event Type
Malfunction
Date Received
May 31, 2013
Date of Event
April 1, 2013
Report Date
June 27, 2013
Manufacturer
ROCHE DIAGNOSTICS
Product Code
GJS
PMA / PMN Number
K062925
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

IT WAS UNKNOWN IF THE INITIAL REPORTER SENT REPORT TO THE FDA.

Description of Event or Problem · 1

CALLER REPORTS THAT DURING A CORRELATION STUDY THE FOLLOWING COAGUCHEK XS/LABORATORY RESULTS WERE OBTAINED: 4.5 INR/2.8 INR, 6.1 INR/4.2 INR, 4.2 INR/3.1 INR, 4.5 INR/3.4 INR, 4.8 INR/3.5 INR. NO ACTIONS TAKEN BASED ON DEVICE RESULTS. NO ADVERSE EVENT REPORTED. REQUESTED RETURN OF SUSPECT SYSTEM AND REPLACEMENT WAS SENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
242151 COAGUCHEK ® XS SYSTEM PROTHROMBIN TIME TEST STRIPS GJS ROCHE DIAGNOSTICS NA 21637312

Patients

Seq Age Sex Outcome Treatment
1