FDA Adverse Event
Malfunction
Summary report: N
LIFEPAK(R) 12 DEFIBRILLATOR/MONITOR SERIES
MDR report key: 3140293
·
Received May 31, 2013
Report
- Report Number
- 3015876-2013-00458
- Event Type
- Malfunction
- Date Received
- May 31, 2013
- Date of Event
- May 4, 2013
- Report Date
- May 4, 2013
- Manufacturer
- PHYSIO-CONTROL, INC
- Product Code
- MKJ
- PMA / PMN Number
- K102972
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
(B)(4): PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE. PHYSIO REPLACED THE POWER PCB ASSEMBLY AND OBSERVED PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS THEN RETURNED TO THE CUSTOMER FOR USE.PHYSIO-CONTROL FURTHER EVALUATED THE REMOVED POWER PCB ASSEMBLY AND DETERMINED THAT THE CAUSE OF THE REPORTED FAILURE WAS DUE TO THE FAILURE OF AN INTEGRATED CIRCUIT CHIP, DESIGNATOR U5.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THEIR DEVICE WOULD NOT POWER ON WITH DC POWER OR THROUGH THE AC POWER ADAPTER. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED FAILURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 242067 | LIFEPAK(R) 12 DEFIBRILLATOR/MONITOR SERIES | DEFIBRILLATORS, AUTOMATIC, EXTERNAL | MKJ | PHYSIO-CONTROL, INC |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |