FDA Adverse Event Malfunction Summary report: N

LIFEPAK(R) 12 DEFIBRILLATOR/MONITOR SERIES

MDR report key: 3140293 · Received May 31, 2013

Report

Report Number
3015876-2013-00458
Event Type
Malfunction
Date Received
May 31, 2013
Date of Event
May 4, 2013
Report Date
May 4, 2013
Manufacturer
PHYSIO-CONTROL, INC
Product Code
MKJ
PMA / PMN Number
K102972
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4): PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE. PHYSIO REPLACED THE POWER PCB ASSEMBLY AND OBSERVED PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS THEN RETURNED TO THE CUSTOMER FOR USE.PHYSIO-CONTROL FURTHER EVALUATED THE REMOVED POWER PCB ASSEMBLY AND DETERMINED THAT THE CAUSE OF THE REPORTED FAILURE WAS DUE TO THE FAILURE OF AN INTEGRATED CIRCUIT CHIP, DESIGNATOR U5.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THEIR DEVICE WOULD NOT POWER ON WITH DC POWER OR THROUGH THE AC POWER ADAPTER. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED FAILURE.