ENDURANT
Report
- Report Number
- 2953200-2013-01038
- Event Type
- Injury
- Date Received
- May 31, 2013
- Report Date
- May 7, 2013
- Manufacturer
- MEDTRONIC IRELAND
- Product Code
- MIH
- PMA / PMN Number
- P100021
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). RESULTS: INHERENT RISK OF PROCEDURE (STENT GRAFT OCCLUSION); PATIENT¿S CONDITION AFFECTED EFFECTIVENESS OF DEVICE (MEDICATION CHANGE); CAUSED BY ANOTHER DRUG/DEVICE (MEDICATION CHANGE). CONCLUSION: KNOWN INHERENT RISK OF PROCEDURE (STENT GRAFT OCCLUSION); DEVICE FAILURE RELATED TO PATIENT CONDITION (MEDICATION CHANGE).
AN ENDURANT STENT GRAFT SYSTEM WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF A 47.3 MM IN DIAMETER ABDOMINAL AORTIC A NEURYSM. THE PROXIMAL NECK WAS 23.1 MM IN DIAMETER. IT WAS REPORTED THAT ON AN UNKNOWN DATE (APPROXIMATELY 6 MONTHS AGO), THE PATIENT WAS HOSPITALIZED FOR HEPATITIS AND THE PATIENT¿S ANTI-COAGULANT THERAPY WAS DISCONTINUED DURING THAT ADMISSION. THE PATIENT CURRENTLY PRESENTED WITH CLAUDICATION. A STUDY DONE IN THE PHYSICIAN OFFICE DETERMINED THE LEFT LIMB OF THE STENT GRAFT TO BE OCCLUDED ( CONTRALATERAL LIMB). THE PHYSICIAN DID NOT FEEL THAT THIS WAS LIMB THREATENING. THE PHYSICIAN DECIDED TO TREAT THE PATIENT WITH A FEMORAL-FEMORAL BYPASS SINCE THE OCCLUSION IS LIKELY CHRONIC, SINCE THE TIME ANTICOAGULATION WAS STOPPED 6 MONTHS AGO. THE PATIENT HAS NOT BEEN TREATED TO DATE AND THE PROCEDURE HAS NOT BEEN SCHEDULED. NO CLINICAL SEQUELAE WERE REPORTED AND THE PATIENT IS FINE. A REVIEW OF A DRAWING SHOWED THAT THE LEFT LIMB WAS COMPLETELY OCCLUDED BEGINNING AT THE FLOW DIVIDER. THE CURRENT AAA SIZE IS 5.1CM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 242357 | ENDURANT | SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT | MIH | MEDTRONIC IRELAND | V01075361 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |