ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-07439
- Event Type
- Malfunction
- Date Received
- May 31, 2013
- Date of Event
- May 4, 2013
- Report Date
- May 4, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS BEEN RETURNED TO ANIMAS BUT EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME.
THE DEVICE WAS RETURNED TO ANIMAS AND EVALUATED BY PRODUCT ANALYSIS ON 07/03/2013 WITH THE FOLLOWING RESULTS: TESTING FOUND AN AUDIO BOLUS BUTTON LEAK FOUND DURING THE LEAK TEST. REMOVED THE PUMP COVER; INTERNAL MOISTURE OBSERVED IN THE PUMP. PUMP POWERS ON TO THE DISPLAY SCREEN WITH VIBRATORY AND AUDIBLE SOUNDS. DISPLAY SCREEN HAS NORMAL CONTRAST. AUDIO BOLUS BUTTON LEAK FOUND DURING. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THE REPORTER CONTACTED ANIMAS ON (B)(6) 2013 ALLEGING THE PUMP EMITTED A CALL SERVICE ALARM AND THEN THE DISPLAY SCREEN WENT BLANK AND THE PUMP WAS VIBRATING AND THE PUMP HAD STOPPED DELIVERING. REMOVING THE BATTERY AND REINSERTING IT SUCCESSFULLY POWERED THE PUMP BACK ON AND NAVIGATION THROUGHOUT THE PUMP WAS POSSIBLE. THE REPORTER STATED THAT PRIOR TO THIS ISSUE OCCURRING, THE PATIENT HAD BEEN WEARING THE PUMP IN THE SPRINKLER. THERE WAS NO REPORTED ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE DELIVERY INTERRUPTION ISSUE REMAINED UNRESOLVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 241862 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 8 YR |