FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3140131 · Received May 31, 2013

Report

Report Number
2531779-2013-07439
Event Type
Malfunction
Date Received
May 31, 2013
Date of Event
May 4, 2013
Report Date
May 4, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS BUT EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO ANIMAS AND EVALUATED BY PRODUCT ANALYSIS ON 07/03/2013 WITH THE FOLLOWING RESULTS: TESTING FOUND AN AUDIO BOLUS BUTTON LEAK FOUND DURING THE LEAK TEST. REMOVED THE PUMP COVER; INTERNAL MOISTURE OBSERVED IN THE PUMP. PUMP POWERS ON TO THE DISPLAY SCREEN WITH VIBRATORY AND AUDIBLE SOUNDS. DISPLAY SCREEN HAS NORMAL CONTRAST. AUDIO BOLUS BUTTON LEAK FOUND DURING. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

THE REPORTER CONTACTED ANIMAS ON (B)(6) 2013 ALLEGING THE PUMP EMITTED A CALL SERVICE ALARM AND THEN THE DISPLAY SCREEN WENT BLANK AND THE PUMP WAS VIBRATING AND THE PUMP HAD STOPPED DELIVERING. REMOVING THE BATTERY AND REINSERTING IT SUCCESSFULLY POWERED THE PUMP BACK ON AND NAVIGATION THROUGHOUT THE PUMP WAS POSSIBLE. THE REPORTER STATED THAT PRIOR TO THIS ISSUE OCCURRING, THE PATIENT HAD BEEN WEARING THE PUMP IN THE SPRINKLER. THERE WAS NO REPORTED ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE DELIVERY INTERRUPTION ISSUE REMAINED UNRESOLVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
241862 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 8 YR