FDA Adverse Event Malfunction Summary report: N

TRILOGY 100

MDR report key: 3140105 · Received May 29, 2013

Report

Report Number
2518422-2013-01004
Event Type
Malfunction
Date Received
May 29, 2013
Date of Event
January 4, 2011
Report Date
January 4, 2011
Manufacturer
RESPIRONICS, INC.
Product Code
CBK
PMA / PMN Number
K083526
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE VENTILATOR WAS RETURNED TO THE MFR FOR EVAL AND THE CUSTOMER'S COMPLAINT WAS CONFIRMED. THE DEVICE'S AC INLET CONNECTOR WAS REPLACED TO ADDRESS THE ISSUE. THIS REPORT IS BEING SUBMITTED AS PART OF A PROGRAM TO RETROSPECTIVELY REVIEW POSSIBLE DEVICE MALFUNCTIONS THAT DID NOT INVOLVE ANY INJURY, TO DETERMINE WHETHER THEY ARE REPORTABLE UNDER 21 CFR PART 803 AND THE MFR'S UPDATED REPORTING PROCEDURES. THIS PROGRAM IS BEING UNDERTAKEN TO ADDRESS FDA WARNING LETTER (B)(4).

Description of Event or Problem · 1

THE MFR REC'D INFO ALLEGED A VENTILATOR'S AC INLET CONNECTOR WAS DAMAGED. THE VENTILATOR WAS NOT IN PT USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
234826 TRILOGY 100 VENTILATOR, CONTINUOUS, FACILITY USE CBK RESPIRONICS, INC. 1054260

Patients

Seq Age Sex Outcome Treatment
1