FDA Adverse Event Malfunction Summary report: N

TRILOGY 200

MDR report key: 3140088 · Received May 29, 2013

Report

Report Number
2518422-2013-00953
Event Type
Malfunction
Date Received
May 29, 2013
Date of Event
January 4, 2011
Report Date
January 4, 2011
Manufacturer
RESPIRONICS, INC.
Product Code
CBK
PMA / PMN Number
K093416
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE VENTILATOR WAS RETURNED TO THE MFR FOR EVAL AND CUSTOMER'S COMPLAINT WAS CONFIRMED. THE DEVICE'S POWER SUPPLY WAS REPLACED TO ADDRESS THE ISSUE. THE MALFUNCTION OF THE POWER SUPPLY WOULD RESULT IN THE FAILURE OF THE DEVICE TO OPERATE ON AC OR THE CHARGE ITS INTERNAL BATTERY. THIS PARTICULAR MALFUNCTION FOR THE POWER SUPPLY HAS ONLY BEEN REPORTED ON DEVICES THAT ARE BEING SOLD INTERNATIONALLY FOR USE WITH 220 VOLT POWER SOURCES. THIS REPORT IS BEING SUBMITTED AS PART OF A PROGRAM TO RETROSPECTIVELY REVIEW POSSIBLE DEVICE MALFUNCTIONS THAT DID NOT INVOLVE ANY INJURY, TO DETERMINE WHETHER THEY ARE REPORTABLE UNDER 21 CFR PART 803 AND THE MFR'S UPDATED REPORTING PROCEDURES. THIS PROGRAM IS BEING UNDERTAKEN TO ADDRESS FDA WARNING LETTER 12-PHI-01.

Description of Event or Problem · 1

A VENTILATOR WAS RETURNED TO THE MFR ALLEGING NO POWER. THERE WAS NO HARM OR INJURY REPORTED.