FDA Adverse Event Malfunction Summary report: N

TRILOGY O2

MDR report key: 3140058 · Received May 29, 2013

Report

Report Number
2518422-2013-01001
Event Type
Malfunction
Date Received
May 29, 2013
Date of Event
April 13, 2011
Report Date
April 13, 2013
Manufacturer
RESPIRONICS, INC.
Product Code
CBK
PMA / PMN Number
K093905
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

DURING THE EVAL OF THE DEVICE AT THE MFR'S SERVICE CENTER, A VENTILATOR INOPERATIVE CODE WAS LOGGED IN THE DEVICE'S ERROR LOG. THE DEVICE'S BLOWER MOTOR WAS REPLACED TO ADDRESS THE ISSUE. THIS REPORT IS BEING SUBMITTED AS PART OF A PROGRAM TO RETROSPECTIVELY REVIEW POSSIBLE DEVICE MALFUNCTIONS THAT DID NOT INVOLVE ANY INJURY, TO DETERMINE WHETHER THEY ARE REPORTABLE UNDER 21 CFR PART 803 AND THE MFR'S UPDATED REPORTING PROCEDURES. THIS PROGRAM IS BEING UNDERTAKEN TO ADDRESS FDA WARNING LETTER (B)(4).

Description of Event or Problem · 1

THE MFR REC'D INFO ALLEGING A "SERVICE REQUIRED" ALARM CONDITION OCCURRED. THE DEVICE WAS NOT IN PT USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
234541 TRILOGY O2 VENTILATOR, CONTINUOUS, FACILITY USE CBK RESPIRONICS, INC. 1040004

Patients

Seq Age Sex Outcome Treatment
1