FDA Adverse Event Injury Summary report: N

PINNACLE MTL INS NEUT36IDX54OD

MDR report key: 3137462 · Received May 30, 2013

Report

Report Number
1818910-2013-17856
Event Type
Injury
Date Received
May 30, 2013
Date of Event
January 5, 2016
Report Date
September 24, 2013
Manufacturer
DEPUY ORTHOPAEDICS INC US
Product Code
KWA
PMA / PMN Number
PK003523
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
ATTORNEY

Narratives

Additional Manufacturer Narrative · 1

UPDATE: (B)(4) 2013 PFS WAS RECEIVED FROM LEGAL, MEDICAL RECORDS WERE RECEIVED FROM LEGAL, AND PART/LOT INFORMATION WAS IDENTIFIED. RECORDS INDICATE THAT THE RIGHT SIDE WAS REVISED BECAUSE OF METALLOSIS AND CORROSION. RECORDS ARE AVAILABLE FOR FURTHER REVIEW. THE INVESTIGATION IS ONGOING. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

ADDITIONAL NARRATIVE: IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. ADDITIONAL INFORMATION: DEPUY CONSIDERS THE INVESTIGATION COMPLETED. SHOULD FURTHER INFORMATION BE RECEIVED, THE COMPLAINT WILL BE UPDATED AT THAT TIME.

Additional Manufacturer Narrative · 1

(B)(4). EXAMINATION OF THE REPORTED DEVICES WAS NOT POSSIBLE AS THEY WERE NOT RETURNED. A SEARCH OF THE COMPLAINTS DATABASES IDENTIFIED OTHER REPORTS AGAINST THE FEMORAL HEAD AND LINER. PER PROCEDURE, THESE DEVICES ARE EXEMPT FROM DEVICE HISTORY RECORD REVIEW. A SEARCH OF THE COMPLAINTS DATABASES IDENTIFIED NO OTHER REPORTS AGAINST THE REMAINING PRODUCT/LOT CODE COMBINATIONS. THE INVESTIGATION CAN DRAW NO CONCLUSIONS WITH THE INFORMATION PROVIDED. BASED ON THE INABILITY TO DETERMINE ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION HAS NOT BEEN INDICATED. DEPUY STILL CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.

Additional Manufacturer Narrative · 1

THE DEVICES ASSOCIATED WITH THIS REPORT WERE NOT RETURNED. A SEARCH OF THE COMPLAINT DATABASE FOUND NO ADDITIONAL RELATED REPORTS FOR THE LOT CODES 1898037, 1914212, 1958202, AND 2125484. A SEARCH OF THE COMPLAINT DATABASE FINDS ADDITIONAL REPORTED INCIDENTS AGAINST LOT CODE 2135037 SINCE ITS RELEASE FOR DISTRIBUTION; HOWEVER, A REVIEW OF THE DEVICE HISTORY RECORDS DID NOT REVEAL ANY RELATED MANUFACTURING ANOMALIES. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY PRODUCT CONTRIBUTION TO THE REPORTED EVENT WITH THE INFORMATION PROVIDED. BASED ON THE INABILITY TO DETERMINE A ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.

Additional Manufacturer Narrative · 1

THE DEVICE ASSOCIATED WITH THIS REPORT WAS NOT RETURNED. A SEARCH OF THE COMPLAINT DATABASE AND/OR DHR REVIEW WAS NOT POSSIBLE AS THE PRODUCT AND LOT CODE REQUIRED WAS NOT PROVIDED. THE INVESTIGATION COULD NOT DRAW ANY CONCLUSIONS REGARDING THE REPORTED EVENT WITH THE INFORMATION AVAILABLE. BASED ON THE INABILITY TO DETERMINE A ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ANY ADDITIONAL INFORMATION BE RECEIVED TO CHANGE THE OUTCOME OF THE PERFORMED INVESTIGATION, THE COMPLAINT WILL BE RE-OPENED.

Description of Event or Problem · 1

LITIGATION ALLEGES DAMAGE TO HIP JOINTS AND BODY.

Description of Event or Problem · 1

UPDATE REC'D 1/5/2016 - PATIENT WAS REVISED DUE TO PAIN AND OSTEOLYTIC LESIONS. CORROSION WAS NOTED ON THE TRUNNION AND INSIDE THE HEAD. THE STEM HAS BEEN ADDED FOR CORROSION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
239059 PINNACLE MTL INS NEUT36IDX54OD HIP LINER KWA DEPUY ORTHOPAEDICS INC US 2135037

Patients

Seq Age Sex Outcome Treatment
1 65 YR Other