FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 3133974
·
Received May 29, 2013
Report
- Report Number
- 2032227-2013-02062
- Event Type
- Malfunction
- Date Received
- May 29, 2013
- Date of Event
- May 10, 2013
- Report Date
- May 10, 2013
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT FAMILY MEMBER OR FRIEND
Narratives
Additional Manufacturer Narrative · 1
A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING, IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE INSULIN PUMP DID NOT DELIVERED THE INSULIN AND THAT THE CUSTOMER HAD HIGH BLOOD GLUCOSE OF 433 MG/DL. CALLER STATED THEY PROGRAMMED 5.0 UNITS, BUT INSULIN IS NOT EXITING. HIGH PRESSURE TEST WAS PERFORMED, AND THE INSULIN PUMP FAILED THE HIGH-PRESSURE TEST TWICE. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 235367 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-722LNAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 80 YR |