COULTER® ACT DIFF 12? ANALYZER

MDR report key: 3133851 · Received May 29, 2013

Report

Report Number
1061932-2013-00980
Event Type
Malfunction
Date Received
May 29, 2013
Date of Event
April 29, 2013
Report Date
April 29, 2013
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K973634
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE FIELD SERVICE ENGINEER (FSE) REPLACED THE PINCH TUBING AT PINCH VALVE VL8 TO CORRECT THE FLUID LEAKING ISSUE. THE FSE PERFORMED A VERIFICATION INSPECTION REPORT (VIP) ON THE INSTRUMENT. THE FSE NOTICED THE DILUENT AND SAMPLE SYRINGE WERE LEAKING AND REPLACED THE DILUENT AND SYRINGE. THE FSE ALSO NOTICED THE PROBE WIPE WAS CORRODED AND REPLACED THE PROBE WIPE. AFTER COMPLETING VIP, THE FSE DISCOVERED PINCH VALVE LV15 WAS NO LONGER FUNCTIONING AND REPLACED THE PINCH VALVE. SERVICE ACTIVITY PERFORMED WAS VERIFIED TO MEET THE SPECIFIED REQUIREMENTS PER ESTABLISHED PROCEDURES. THE INSTRUMENT CONFORMED TO THE MANUFACTURER'S PUBLISHED SPECIFICATIONS AND WAS RETURNED TO NORMAL OPERATION. PROBE WIPE (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED APPROXIMATELY TEN TO TWENTY MILLILITERS OF FLUID LEAKED OUTSIDE THE INSTRUMENT AND ONTO THE COUNTER FROM THE COULTER ACT DIFF 12¿ ANALYZER DURING SYSTEM STARTUP. THE OPERATOR WAS WEARING PROTECTIVE GLOVES AT THE TIME OF THE EVENT AND DID NOT HAVE DIRECT CONTACT WITH THE FLUID. THERE WAS NO OPERATOR INJURY OR ADVERSE EFFECT ASSOCIATED WITH THIS EVENT. PATIENT RESULTS WERE NOT IMPACTED, AND THERE WAS NO PATIENT CONSEQUENCE ASSOCIATED WITH THIS EVENT. THE FACILITY HAS AN EXPOSURE CONTROL OR RISK MANAGEMENT PLAN IN PLACE. A BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO ASSESS THE INSTRUMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
235558 COULTER® ACT DIFF 12? ANALYZER COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER NA NA