FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3133643 · Received May 29, 2013

Report

Report Number
3004209178-2013-94162
Event Type
Malfunction
Date Received
May 29, 2013
Date of Event
May 13, 2013
Report Date
May 13, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH A DETACHED END CAP. THE DRIVE SUPPORT DISK WAS INSPECTED AND NO ANOMALY WAS NOTED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE DRIVE SUPPORT CAP WAS POPPED OUT COMPLETELY, AND IT HAPPENED DURING THE PRIME PROCESS. THE BLOOD GLUCOSE READING WAS 130MG/DL. ADVISED THE CALLER TO DISCONNECT FROM THE INSULIN PUMP AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
237325 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722NAS

Patients

Seq Age Sex Outcome Treatment
1 53 YR