PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 3133483
·
Received May 29, 2013
Report
- Report Number
- 3004209178-2013-94119
- Event Type
- Malfunction
- Date Received
- May 29, 2013
- Date of Event
- January 4, 2013
- Report Date
- May 9, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP ALARMED MOTOR ERROR DURING THE BASIC OCCLUSION TEST, AND THE DEVICE WAS UNABLE TO PRIME DURING THE PRIME TEST AS A RESULT OF A LOOSE DRIVE SUPPORT DISK. THE DEVICE HAD CRACKED CASE NEAR ALL CORNERS OF THE DISPLAY WINDOW.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE CUSTOMER'S INSULIN PUMP ALARMED SEVERAL TIMES. NO FURTHER INFORMATION WAS PROVIDED.