FDA Adverse Event Malfunction Summary report: N

TAXUS? LIBERTÉ?

MDR report key: 3133477 · Received May 29, 2013

Report

Report Number
2134265-2013-03386
Event Type
Malfunction
Date Received
May 29, 2013
Date of Event
February 14, 2013
Report Date
April 30, 2013
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
NIQ
PMA / PMN Number
P060008
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
EG
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATED BY MFR.: RETURNED PRODUCT CONSISTED OF A TAXUS LIBERTE STENT DELIVERY SYSTEM (SDS) WITH STENT AND NO ORIGINAL PACKAGING OR OTHER DEVICES. THERE WAS CONTRAST AND BLOOD UNDER THE BALLOON FOLDS AND BETWEEN THE STENT STRUTS. THE BALLOON WAS TIGHTLY FOLDED WITH THE STENT SECURED ON THE BALLOON. THE FIRST, SECOND, THIRD, SEVENTH, NINTH, ELEVENTH AND THIRTEENTH ROWS OF STRUTS ON THE PROXIMAL END OF THE STENT WERE BENT, FLARED AND OVERLAPPING. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL PROCEDURAL FACTOR. (B)(4).

Description of Event or Problem · 1

THIS COMPLAINT IS NOW REPORTABLE BASED ON THE ANALYSIS COMPLETED ON (B)(4) 2013. IT WAS REPORTED THAT DURING A CORONARY STENTING PROCEDURE A STENT FAILED TO CROSS THE LESION OCCURRED. ACCESS WAS OBTAINED VIA RIGHT FEMORAL ARTERY. THE TARGET LESION WAS LOCATED IN THE RIGHT CORONARY ARTERY (RCA). A 28 X 2.50MM TAXUS LIBERTÉ STENT DELIVERY SYSTEM WAS SELECTED FOR TREATMENT. UPON ADVANCING OF THE COMPLAINT DEVICE, THE STENT FAILED TO CROSS THE LESION AND THE DEVICE WAS REMOVED AS A WHOLE. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATION WAS REPORTED AND PATIENT'S CONDITION WAS STABLE. HOWEVER RETURNED DEVICE ANALYSIS REVEALED STENT DAMAGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
236581 TAXUS? LIBERTÉ? CORONARY DRUG-ELUTING STENT NIQ BOSTON SCIENTIFIC - MAPLE GROVE H7493894028250 15514620

Patients

Seq Age Sex Outcome Treatment
1 52 YR