FDA Adverse Event
Malfunction
Summary report: N
REMB MICRO DRILL
MDR report key: 3133461
·
Received May 29, 2013
Report
- Report Number
- 0001811755-2013-01222
- Event Type
- Malfunction
- Date Received
- May 29, 2013
- Date of Event
- April 12, 2013
- Report Date
- April 16, 2013
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- DZJ
- PMA / PMN Number
- K040369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE REPORTED EVENT, HANDPIECE RUN-ON, WAS NOT DUPLICATED; IT WAS CONFIRMED THAT THE DEVICE HAD A CORROSION BY THE SERVICE TECHNICIAN THROUGH FUNCTIONAL EVALUATION, DISASSEMBLY AND VISUAL INSPECTION. UPON DISASSEMBLY, THE MOTOR WAS FOUND CORRODED. THE PRESENCE OF A MOTOR CORROSION IS LIKELY TO CAUSE OR CONTRIBUTE TO RUN-ON.
Description of Event or Problem · 1
THE REMB MICRO DRILL WAS SENT FOR EVALUATION DUE TO RUNNING ON ITS OWN AFTER USE DURING TESTING. THERE WAS NO PATIENT INVOLVEMENT AND NO ADVERSE CONSEQUENCES ASSOCIATED WITH THE EVENT.