FDA Adverse Event Malfunction Summary report: N

SABO SAG SAW

MDR report key: 3133435 · Received May 29, 2013

Report

Report Number
0001811755-2013-01225
Event Type
Malfunction
Date Received
May 29, 2013
Date of Event
April 30, 2013
Report Date
May 1, 2013
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
HRX
PMA / PMN Number
K943323
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE REPORTED EVENT OF THE TIP PIECE FALLING OFF WAS CONFIRMED. SIGNS OF THIRD PARTY WORK ON VARIOUS COMPONENTS OF THE DEVICE WERE NOTED DURING FAILURE ANALYSIS. UNAUTHORIZED MODIFICATION OF THESE COMPONENTS IS A PROBABLE CAUSE OF THE REPORTED DISASSEMBLY. THE DEVICE INSTRUCTIONS FOR USE STATE THAT THE DEVICE SHOULD NOT BE DISASSEMBLED OR ATTEMPTED TO BE SERVICED, AND THAT NO SYSTEM COMPONENTS OR ACCESSORIES SHOULD BE MODIFIED. THE INSTRUCTIONS FOR USE FURTHER STATE THAT ALL CONCERNS SHOULD BE DIRECTED TO STRYKER CUSTOMER SERVICE. STRYKER RECOMMENDS THAT ALL FAILURES AND MAINTENANCE ASSOCIATED WITH ITS DEVICES SHOULD BE DIRECTED TO TRAINED STRYKER SERVICE PROFESSIONALS.

Additional Manufacturer Narrative · 1

(B)(4): FAILURE ANALYSIS IN PROGRESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT PRIOR TO THE START OF A SURGICAL PROCEDURE AT THE USER FACILITY THE TIP FELL OFF THE SAGITTAL HEAD OF THE DEVICE. NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT. THE USER FACILITY REPORTED PREVIOUS THIRD PARTY REPAIR, WHICH WAS CONFIRMED UPON EVALUATION AT THE MANUFACTURER FACILITY.

Description of Event or Problem · 1

IT WAS REPORTED THAT PRIOR TO THE START OF A SURGICAL PROCEDURE AT THE USER FACILITY THE TIP FELL OFF THE SAGITTAL HEAD OF THE DEVICE. NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT. THE USER FACILITY REPORTED PREVIOUS THIRD PARTY REPAIR, WHICH WAS CONFIRMED UPON EVALUATION AT THE MANUFACTURER FACILITY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
234223 SABO SAG SAW ARTHROSCOPE HRX STRYKER INSTRUMENTS-KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1