PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
MDR report key: 3133434
·
Received May 29, 2013
Report
- Report Number
- 3004209178-2013-94052
- Event Type
- Injury
- Date Received
- May 29, 2013
- Date of Event
- April 25, 2013
- Report Date
- April 29, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP HAD INTERMITTENT BUTTON RESPONSE AS A RESULT OF A CORRODED KEYPAD TRACE. UNABLE TO CONFIRM BUTTON ERROR ALARMS.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE INSULIN PUMP ALARMED BUTTON ERROR. THE BLOOD GLUCOSE READING WAS 300MG/DL. TROUBLESHOOTING WAS PERFORMED. THE CUSTOMER MENTIONED BEING HOSPITALIZED AND NOT SURE IF IT WAS DIABETES RELATED OR NOT. ADVISED THE CALLER THAT THE DEVICE WOULD BE REPLACED AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 236013 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723NAP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 49 YR | Hospitalization |