FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3133421 · Received May 29, 2013

Report

Report Number
3004209178-2013-94100
Event Type
Malfunction
Date Received
May 29, 2013
Date of Event
February 9, 2013
Report Date
May 9, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE DEVICE WAS RECEIVED WITH CURRENTS WITHIN SPECIFICATIONS. HOWEVER, THE INSULIN PUMP ALARMED MOTOR ERROR DURING REWIND AS A RESULT OF A CORRODED MOTOR HOME SWITCH. THE DEVICE WAS RETURNED WITH PROTRUDED/LOOSE DRIVE SUPPORT DISK.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR, AND SHE WAS NOT ABLE TO REWIND THE DEVICE. THE BLOOD GLUCOSE READING WAS 395MG/DL. TROUBLESHOOTING WAS PERFORMED, AND THE DRIVE SUPPORT CAP HAS NO DAMAGE. ADVISED THE CALLER TO DISCONTINUE THE USE OF THE INSULIN PUMP AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
237438 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722NAS

Patients

Seq Age Sex Outcome Treatment
1 51 YR