FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 3133421
·
Received May 29, 2013
Report
- Report Number
- 3004209178-2013-94100
- Event Type
- Malfunction
- Date Received
- May 29, 2013
- Date of Event
- February 9, 2013
- Report Date
- May 9, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS REVEALED THAT THE DEVICE WAS RECEIVED WITH CURRENTS WITHIN SPECIFICATIONS. HOWEVER, THE INSULIN PUMP ALARMED MOTOR ERROR DURING REWIND AS A RESULT OF A CORRODED MOTOR HOME SWITCH. THE DEVICE WAS RETURNED WITH PROTRUDED/LOOSE DRIVE SUPPORT DISK.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR, AND SHE WAS NOT ABLE TO REWIND THE DEVICE. THE BLOOD GLUCOSE READING WAS 395MG/DL. TROUBLESHOOTING WAS PERFORMED, AND THE DRIVE SUPPORT CAP HAS NO DAMAGE. ADVISED THE CALLER TO DISCONTINUE THE USE OF THE INSULIN PUMP AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 237438 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-722NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 51 YR |