MEGA SUTURECUT NEEDLE DRIVER INSTRUMENT
Report
- Report Number
- 2955842-2013-01863
- Event Type
- Malfunction
- Date Received
- May 29, 2013
- Date of Event
- May 8, 2013
- Report Date
- May 9, 2013
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI
- Reporter Occupation
- OTHER
Narratives
INTUITIVE SURGICAL, INC. RECEIVED THE INSTRUMENT INVOLVED WITH THE COMPLAINT. ENGINEERING CONFIRMED A BROKEN GRIP CLOSE CABLE AT THE DISTAL IDLERS. THE IDLER PULLEY SPUN FREELY AND WAS NOT DAMAGED. THE CABLE SEGMENT WAS STICKING OUT AT THE INSTRUMENT'S WRIST. OTHER CABLES AT THE INSTRUMENT'S WRIST WERE NOT DAMAGED. ADDITIONAL OBSERVATIONS THAT WERE NOT INITIALLY REPORTED BY THE CUSTOMER WERE SCRATCHES ON THE SURFACE OF THE DISTAL PULLEY. ENGINEERING WAS UNABLE TO CONCLUDE HOW THE PULLEY DAMAGE OCCURRED. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO REOCCUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.
DURING A DA VINCI S HYSTERECTOMY PROCEDURE, THE CABLE ON THE MEGA SUTURECUT NEEDLE DRIVER INSTRUMENT REPORTEDLY BROKE. NOTHING REPORTEDLY FELL INTO A PATIENT. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 237307 | MEGA SUTURECUT NEEDLE DRIVER INSTRUMENT | ENDOSCOPIC INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420309-01 | M10130211 720 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DA VINCI SI SYSTEM INSTRUMENTS AND ACCESSORIES |