FDA Adverse Event Malfunction Summary report: N

MEGA SUTURECUT NEEDLE DRIVER INSTRUMENT

MDR report key: 3133400 · Received May 29, 2013

Report

Report Number
2955842-2013-01863
Event Type
Malfunction
Date Received
May 29, 2013
Date of Event
May 8, 2013
Report Date
May 9, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INTUITIVE SURGICAL, INC. RECEIVED THE INSTRUMENT INVOLVED WITH THE COMPLAINT. ENGINEERING CONFIRMED A BROKEN GRIP CLOSE CABLE AT THE DISTAL IDLERS. THE IDLER PULLEY SPUN FREELY AND WAS NOT DAMAGED. THE CABLE SEGMENT WAS STICKING OUT AT THE INSTRUMENT'S WRIST. OTHER CABLES AT THE INSTRUMENT'S WRIST WERE NOT DAMAGED. ADDITIONAL OBSERVATIONS THAT WERE NOT INITIALLY REPORTED BY THE CUSTOMER WERE SCRATCHES ON THE SURFACE OF THE DISTAL PULLEY. ENGINEERING WAS UNABLE TO CONCLUDE HOW THE PULLEY DAMAGE OCCURRED. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO REOCCUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

DURING A DA VINCI S HYSTERECTOMY PROCEDURE, THE CABLE ON THE MEGA SUTURECUT NEEDLE DRIVER INSTRUMENT REPORTEDLY BROKE. NOTHING REPORTEDLY FELL INTO A PATIENT. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
237307 MEGA SUTURECUT NEEDLE DRIVER INSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420309-01 M10130211 720

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SYSTEM INSTRUMENTS AND ACCESSORIES