FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3133234 · Received May 29, 2013

Report

Report Number
2531779-2013-07150
Event Type
Malfunction
Date Received
May 29, 2013
Report Date
May 1, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: THE PUMP DISPLAY SCREEN WAS FOUND TO HAVE VERTICAL LINES THROUGH THE LETTERING. THE PUMP WAS OPENED AND THE DISPLAY SCREEN WAS RESEATED WITHOUT RESOLVING THE LINES THROUGH THE DISPLAY. THE PUMP DISPLAY WAS REPLACED WITH A TEST DISPLAY AND THE ISSUE RESOLVED. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

THE PUMP WAS RETURNED TO ANIMAS. THE INVESTIGATION REVEALED A FAILED DISPLAY FLEX. THIS REPORT IS MADE BASED ON THE RESULTS OF INVESTIGATION.