FDA Adverse Event Malfunction Summary report: N

ABSOLUTE PRO VASCULAR SELF-EXPANDING STENT SYSTEM

MDR report key: 3133229 · Received May 29, 2013

Report

Report Number
2024168-2013-03313
Event Type
Malfunction
Date Received
May 29, 2013
Date of Event
May 3, 2013
Report Date
May 3, 2013
Manufacturer
AV-TEMECULA-CT
Product Code
NIO
PMA / PMN Number
P110028
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR ANALYSIS. THE PREMATURE DEPLOYMENT WAS ABLE TO BE CONFIRMED. BASED ON A VISUAL AND FUNCTIONAL INSPECTION OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. THE RESULTS OF THE QUERY OF SIMILAR INCIDENTS IN THE COMPLAINT HANDLING DATABASE FOR THIS LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE REVIEWED INFORMATION, NO PRODUCT DEFICIENCY WAS IDENTIFIED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING PREPARATION OF THE 10/40 MM ABSOLUTE PRO STENT SYSTEM, ON THE BACK TABLE, THE PHYSICIAN REMOVED THE STYLET; HOWEVER, THE PROTECTIVE SHEATH WAS INADVERTENTLY PULLED, AND THE STENT BECAME PARTIALLY DEPLOYED. THE DEVICE WAS NOT USED IN THE ANATOMY, AND A NEW ABSOLUTE PRO STENT SYSTEM WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED. IT WAS NOTED THAT THE STENT WAS COMPLETELY DEPLOYED DURING PACKAGING FOR RETURN, AND THE STENT WAS DISCARDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
235710 ABSOLUTE PRO VASCULAR SELF-EXPANDING STENT SYSTEM SELF EXPANDING STENT SYSTEM NIO AV-TEMECULA-CT 2050361

Patients

Seq Age Sex Outcome Treatment
1