FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 3132943 · Received May 29, 2013

Report

Report Number
1416980-2013-13559
Event Type
Malfunction
Date Received
May 29, 2013
Date of Event
March 19, 2013
Report Date
May 7, 2013
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER, AND THE EVALUATION IS COMPLETE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT. THE REPORTED DIFFICULTY OF AN IIPV EVENT WAS CONFIRMED THROUGH AN EVENT HISTORY LOG REVIEW. THE CAUSE WAS DETERMINED TO BE FALSE EMPTY DETECT AND USE ERROR, CLINICIAN INAPPROPRIATELY SET MINIMUM DRAIN VOLUME PERCENT SETTING TOO LOW. (B)(4). SECTION 12 "PROGRAMMING A PRESCRIPTION" IN 12.3.3 ON PAGE 12-6 GIVES THE WARNING THAT "IF YOU SET THE MINIMUM DRAIN VOLUME% TOO LOW, AN INCOMPLETE DRAIN COULD RESULT FOR THE PATIENT. THIS COULD CAUSE AN INCREASED INTRAPERITONEAL VOLUME (IIPV) SITUATION." ALSO, PAGE 12-7 IN TABLE 2-2 NURSE'S MENU SETTINGS, STATES "THE DEFAULT SETTING OF 85% IS APPROPRIATE FOR PATIENTS WITH LOW TO MODERATE ULTRAFILTRATION RATES (UF PER CYCLE IS LESS THAN OR EQUAL TO 10% OF THE FILL VOLUME). A HIGHER SETTING MAY BE MORE APPROPRIATE FOR PATIENTS WITH HIGHER ULTRAFILTRATION RATES (UF PER CYCLE IS GREATER THAN 10% OF THE FILL VOLUME). IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP WILL BE SENT.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE MACHINE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT(S) WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2013 21:13:09. DURING NIGHT DRAIN CYCLE FOUR, THE PATIENT'S ULTRAFILTRATION READING WAS 1744ML, INDICATING THE HOME PATIENT (HP) DRAINED 1744ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 2000ML THIS INFORMATION MEETS IIPV CRITERIA."NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
235519 HOMECHOICE PRO SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1 77 YR