ENSEAL G2 STRAIGHT JAW
Report
- Report Number
- 3005075853-2013-02598
- Event Type
- Malfunction
- Date Received
- May 29, 2013
- Date of Event
- May 8, 2013
- Report Date
- May 9, 2013
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GEI
- PMA / PMN Number
- K112033
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). SHOULD THE INFORMATION BE PROVIDED LATER, A SUPPLEMENTAL MEDWATCH WILL BE SENT. THE DEVICE WAS RECEIVED WITH THE TIP OF I BLADE MISSING AND NOT RETURNED. IN ADDITION THE UPPER JAWS DETACHED FROM INSTRUMENT AND RETURNED FOR ANALYSIS. THE DEVICE WAS PARTIALLY TESTED WITH A GENERATOR AND THE DEVICE PERFORMED AS REQUIRED DURING TESTING; THOUGH ALL TESTING COULD NOT BE COMPLETED DUE TO THE DAMAGED I BLADE AND JAW MISSING. THERE IS INSUFFICIENT EVIDENCE RELATED TO WHAT CAUSED THE DAMAGE; A PROBABLE CAUSE OF THE DAMAGE TO THE I-BLADE COULD BE NOT ALLOWING THICK AND FIBROUS TISSUES TO DENATURE PRIOR TO I-BLADE ADVANCEMENT.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC GASTRIC BYPASS PROCEDURE THERE WERE ADHESIONS FROM A PREVIOUS OPEN SURGERY PRESENT AT THE SITE. THE SURGEON WAS TAKING DOWN THE ADHESIONS AND THE JAW FELL OFF INTO THE PATIENT. THE BLADE HAD BROKEN IN TWO PIECES; THEY REMOVED BOTH OF THE PIECES FROM THE PATIENT. THEY USED AN NSEAL545RH TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE REPORTED. THE DEVICE IS RETURNING FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 235514 | ENSEAL G2 STRAIGHT JAW | ELECTROSURGICAL CUTTING & COAGULATION & ACCESSORIES | GEI | ETHICON ENDO-SURGERY, LLC. | UNK | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | GENERATOR |