FDA Adverse Event Malfunction Summary report: N

ENSEAL G2 STRAIGHT JAW

MDR report key: 3132928 · Received May 29, 2013

Report

Report Number
3005075853-2013-02598
Event Type
Malfunction
Date Received
May 29, 2013
Date of Event
May 8, 2013
Report Date
May 9, 2013
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GEI
PMA / PMN Number
K112033
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). SHOULD THE INFORMATION BE PROVIDED LATER, A SUPPLEMENTAL MEDWATCH WILL BE SENT. THE DEVICE WAS RECEIVED WITH THE TIP OF I BLADE MISSING AND NOT RETURNED. IN ADDITION THE UPPER JAWS DETACHED FROM INSTRUMENT AND RETURNED FOR ANALYSIS. THE DEVICE WAS PARTIALLY TESTED WITH A GENERATOR AND THE DEVICE PERFORMED AS REQUIRED DURING TESTING; THOUGH ALL TESTING COULD NOT BE COMPLETED DUE TO THE DAMAGED I BLADE AND JAW MISSING. THERE IS INSUFFICIENT EVIDENCE RELATED TO WHAT CAUSED THE DAMAGE; A PROBABLE CAUSE OF THE DAMAGE TO THE I-BLADE COULD BE NOT ALLOWING THICK AND FIBROUS TISSUES TO DENATURE PRIOR TO I-BLADE ADVANCEMENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC GASTRIC BYPASS PROCEDURE THERE WERE ADHESIONS FROM A PREVIOUS OPEN SURGERY PRESENT AT THE SITE. THE SURGEON WAS TAKING DOWN THE ADHESIONS AND THE JAW FELL OFF INTO THE PATIENT. THE BLADE HAD BROKEN IN TWO PIECES; THEY REMOVED BOTH OF THE PIECES FROM THE PATIENT. THEY USED AN NSEAL545RH TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE REPORTED. THE DEVICE IS RETURNING FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
235514 ENSEAL G2 STRAIGHT JAW ELECTROSURGICAL CUTTING & COAGULATION & ACCESSORIES GEI ETHICON ENDO-SURGERY, LLC. UNK ASKU

Patients

Seq Age Sex Outcome Treatment
1 GENERATOR