FDA Adverse Event Malfunction Summary report: N

GYNECARE MORCELLEX* TISSUE MORCELLATOR

MDR report key: 3132889 · Received May 29, 2013

Report

Report Number
2210968-2013-05992
Event Type
Malfunction
Date Received
May 29, 2013
Date of Event
May 6, 2013
Report Date
May 6, 2013
Manufacturer
ETHICON, INC.
Product Code
HET
PMA / PMN Number
K100280
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). POOR BLADE PERFORMANCE. THE DEVICE WORKED INTERMITTENTLY AND PERFORMED POORLY, CUTTING SLOWLY AND THE CORD SHAKING. THIS IS ONE OF FOUR MEDWATCHES BEING SUBMITTED AS FOUR DEVICES WERE INVOLVED IN THIS EVENT. SEE ALSO MEDWATCHES 2210968-2013-06484, 2210968-2013-06492, AND 2210968-2013-05993. THE SAME PATIENT IS REPRESENTED IN EACH MEDWATCH.

Additional Manufacturer Narrative · 1

(B)(4). CONCLUSION: THE PRODUCT UPON WHICH THIS MEDWATCH IS BASED IS ANTICIPATED. ONCE THE PRODUCT IS RECEIVED, ANY FURTHER INFORMATION DERIVED FROM THE EVALUATION WILL BE SUBMITTED IN A SUPPLEMENTAL 3500A FORM. IN ADDITION, A REVIEW OF THE BATCH MANUFACTURING RECORDS WAS CONDUCTED AND THE BATCH MET ALL FINISHED GOODS RELEASE CRITERIA. THIS IS ONE OF TWO MEDWATCHES BEING SUBMITTED AS TWO DEVICES WERE INVOLVED IN THIS EVENT. SEE ALSO MEDWATCH 2210968-2013-05993. THE SAME PATIENT IS REPRESENTED IN EACH MEDWATCH.

Additional Manufacturer Narrative · 1

(B)(4): THE ACTUAL TRIGGER OF THE DEVICE INVOLVED IN THIS EVENT WAS RETURNED FOR EVALUATION. THE MAIN HOUSING WAS NOT RETURNED. THE DEVICE WAS VISUALLY AND FUNCTIONALLY EVALUATED. UPON EVALUATION, THE TRIGGER OPERATED AS INTENDED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PATIENT UNDERWENT A HYSTERECTOMY ON (B)(6) 2013. DURING THE PROCEDURE, THE CUTTING ACTION OF THE DEVICE WAS POOR. ANOTHER LIKE DEVICE WAS USED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
236482 GYNECARE MORCELLEX* TISSUE MORCELLATOR LAPAROSCOPE, GYNECOLOGIC HET ETHICON, INC. NA MT217497

Patients

Seq Age Sex Outcome Treatment
1