FDA Adverse Event Malfunction Summary report: N

MONOPOLAR CURVED SCISSORS INSTRUMENT

MDR report key: 3132866 · Received May 29, 2013

Report

Report Number
2955842-2013-01858
Event Type
Malfunction
Date Received
May 29, 2013
Date of Event
May 8, 2013
Report Date
May 10, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. ENGINEERING EVALUATION CONFIRMED THAT THE TUBE EXTENSION (ORANGE TAPE) WAS BROKEN AND MISSING A PIECE AT THE DISTAL END OF THE INSTRUMENT. THE CLEVIS WAS NOT DISLODGED FROM TUBE EXTENSION. THE TUBE EXTENSION LIKELY BROKE DUE TO EXCESSIVE SIDE LOADING OR OTHER TYPE OF MISHANDLING. ELECTRICAL CONTINUITY TESTING WAS PERFORMED AND PASSED. THE INSTRUMENTS & ACCESSORIES INSTRUCTIONS FOR USE (IFU) SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO REOCCUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

PRIOR TO STARTING A DA VINCI SI SURGICAL PROCEDURE, THE USER FACILITY IDENTIFIED THE ORANGE TAPE PEELING BACK BY THE TIP OF THE MONOPOLAR CURVED SCISSORS INSTRUMENT. NO MISSING OR FALLEN PIECES WERE REPORTED. THE INSTRUMENT REPORTEDLY WAS NOT USED ON A PATIENT AFTER THE REPORTED ISSUE WAS IDENTIFIED AND THERE WAS NO ALLEGATION OF HARM OR INJURY TO A PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
236297 MONOPOLAR CURVED SCISSORS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420179-10 M11130322 606

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SYSTEM INSTRUMENTS, ACCESSORIES, ESU