HOMECHOICE PRO
Report
- Report Number
- 1416980-2013-13553
- Event Type
- Malfunction
- Date Received
- May 29, 2013
- Date of Event
- March 13, 2013
- Report Date
- May 7, 2013
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- Removal / Correction Number
- 1423500-01/08/10-001-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE WAS RETURNED TO BAXTER, AND THE EVALUATION IS COMPLETE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT OPENED DURING INVESTIGATION OF (B)(4). THE REPORTED DIFFICULTY OF AN IIPV EVENT WAS CONFIRMED THROUGH AN EVENT HISTORY LOG REVIEW. THE ASSIGNABLE CAUSE WAS DETERMINED TO BE FALSE EMPTY DETECT AND USE ERROR, CLINICIAN INAPPROPRIATELY SET MINIMUM DRAIN VOLUME PERCENT SETTING TOO LOW. THE FOLLOWING LABELING WAS REVIEWED: B)(4) 2009 HOMECHOICE APD SYSTEMS TRAINER'S GUIDE. SECTION 12 "NURSE'S SETTINGS" IN 12.3.2 ON PAGE 12-6 GIVES THE WARNING THAT "IF YOU SET THE MINIMUM DRAIN VOLUME% TOO LOW, AN INCOMPLETE DRAIN COULD RESULT FOR YOUR PATIENT. THIS COULD CAUSE AN INCREASED INTRAPERITONEAL VOLUME (IIPV) SITUATION." ALSO, PAGE 12-7 IN TABLE 2-2 NURSE'S MENU SETTINGS, STATES "THE DEFAULT SETTING OF 85% IS APPROPRIATE FOR PATIENTS WITH LOW TO MODERATE ULTRAFILTRATION RATES (UF PER CYCLE IS LESS THAN OR EQUAL TO 10% OF THE FILL VOLUME). A HIGHER SETTING MAY BE MORE APPROPRIATE FOR PATIENTS WITH HIGHER ULTRAFILTRATION RATES (UF PER CYCLE IS GREATER THAN 10% OF THE FILL VOLUME)." IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP WILL BE SENT.
DURING EVALUATION OF A RETURNED HOMECHOICE MACHINE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2013 AT 00:07:27. DURING NIGHT DRAIN CYCLE SIX, THE PATIENT'S ULTRAFILTRATION READING WAS 1696ML, INDICATING THE HOME PATIENT (HP) DRAINED 1696ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 2000ML. THIS INFORMATION MEETS IIPV CRITERIA. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 237315 | HOMECHOICE PRO | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - LARGO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |