FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3132801 · Received May 29, 2013

Report

Report Number
2531779-2013-07094
Event Type
Malfunction
Date Received
May 29, 2013
Report Date
April 30, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) /2013 WITH THE FOLLOWING FINDINGS: INVESTIGATION REVEALED THAT THE AUDIBLE ALERTS WERE NOT FUNCTIONAL. THE VIBRATORY ALERTS WERE FOUND TO BE FUNCTIONING APPROPRIATELY. A DAMAGED SOLDER JOINT WAS OBSERVED DURING INVESTIGATION. UNRELATED TO THE AUDITORY ALERT ISSUE, INVESTIGATION REVEALED STRIPPED BATTERY CAP THREADS AND A CRACKED BATTERY COMPARTMENT. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. (B)(6).

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED THAT THE AUDIBLE ALERTS WERE NOT FUNCTIONAL. THERE WAS NO INDICATION THAT THE DEVICE CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(4) 2013.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
235933 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 25 YR