ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-07140
- Event Type
- Malfunction
- Date Received
- May 29, 2013
- Report Date
- May 1, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- PATIENT
Narratives
SUBMITTED: (B)(4) 2013. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. DEVICE EVALUATION: REVIEW OF THE BLACK BOX REVEALED THE TIME AND DATE WERE INCORRECT THROUGHOUT THE ENTIRE HISTORY. AS A RESULT, THE TOTAL DAILY DOSE HISTORY IN THE PUMP WAS RUNNING FROM THE FACTORY DEFAULT SETTING SINCE 11/19/2012. NO TIME/DATE ISSUE WAS OBSERVED. ON TESTING, THE PUMP POWERED ON PROPERLY TO THE VERIFY SCREEN. THE DISPLAY WAS NOTED TO BE DIM. A TEST DISPLAY WAS ATTACHED TO THE PUMP WITH ILLUMINATED PROPERLY. THE PUMP WAS EXERCISED FOR 24 HOURS WITHOUT ALARMS. THE PUMP WAS POWERED OFF ON (B)(4) 2013 FOR A TOTAL OF 70 MINUTES¿ TIME. WHEN THE PUMP WAS REBOOTED, THE TIME AND DATE HAD NOT BEEN MAINTAINED PROPERLY. THE PUMP WAS OPENED FOR INVESTIGATION AND REVEALED THE INTERNAL BATTERY WAS LEAKING. ADDITIONALLY, THE BOLUS BUTTON COVER WAS NOTED TO BE TORN EXPOSING THE SLUG CONTACT UNDERNEATH. THE AUDIO BOLUS BUTTON WAS RESPONSIVE WHEN TESTED. THE BOLUS BUTTON COVER WAS REMOVED FOR INVESTIGATION AND DID NOT REVEAL ANY EVIDENCE OF CONTAMINATION OF THE CONTACT.
THERE IS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(6) 2013. THE AUDIO BOLUS BUTTON WAS DAMAGED, BUT OPERABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 235116 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 18 YR |